Users Guide
Agilent Technologies
ESA SpectrumAnalyzers
This guide documents firmware revision A.05.xx
Manufacturing Part Number: E4401-90189
Supersedes E4401-90174 Includes changes documented in E4401-90254
Printed in USA
August 2000
© Copyright 2000 Agilent Technologies
This manual provides documentation for the following instruments:
Agilent Technologies ESA-E Series
Agilent Technologies ESA-L Series
and
E4401B (9 kHz–1.5 GHz)
E4402B (9 kHz–3.0 GHz)
E4404B (9 kHz–6.7 GHz)
E4405B (9 kHz–13.2 GHz)
E4407B (9 kHz–26.5 GHz)
E4411B (9 kHz–1.5 GHz)
E4403B (9 kHz–3.0 GHz)
E4408B (9 kHz–26.5 GHz)
Zobrazit stránku 0
1 2 3 4 5 6 ... 320 321

Shrnutí obsahu

Strany 1 - User’s Guide

User’s GuideAgilent TechnologiesESA SpectrumAnalyzersThis guide documents firmware revision A.05.xxManufacturing Part Number: E4401-90189Supersedes E4

Strany 2

xContentsRF Communications Hardware (Option B7E) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10Distance to Fault Accessory Kit (

Strany 3 - LIMITATION OF WARRANTY

3-40 Chapter3Functional TestingScale FidelityScale FidelityTest LimitsTest DescriptionA 50 MHz CW signal is applied from a synthesized sweeper to the

Strany 4 - EXCLUSIVE REMEDIES

Chapter 3 3-41Functional TestingScale FidelityFigure 3-5Procedure1. Connect the equipment as shown in Figure 3-5.2. Preset the synthesized sweeper. Se

Strany 5

3-42 Chapter3Functional TestingScale FidelityPeak Search4. Adjust the amplitude on the sythesized sweeper until the markeramplitude on the analyzer re

Strany 6

Chapter 3 3-43Functional TestingSecond Harmonic Spurious ResponsesSecond Harmonic Spurious ResponsesTest LimitsTest DescriptionTo test second harmonic

Strany 7

3-44 Chapter3Functional TestingSecond Harmonic Spurious ResponsesRequired EquipmentFigure 3-6Instrument Critical Specifications(for this test)Recommend

Strany 8

Chapter 3 3-45Functional TestingSecond Harmonic Spurious ResponsesProcedure1. Connect the equipment as shown in Figure 3-6.2. Set the synthesized swee

Strany 9

3-46 Chapter3Functional TestingTracking Generator Level Flatness: HP E4401B and E4411B, Options 1DNand 1DQTracking Generator Level Flatness:HP E4401B

Strany 10 - Contents

Chapter 3 3-47Functional TestingTracking Generator Level Flatness: HP E4401B and E4411B, Options 1DNand 1DQFigure 3-7Procedure1. Calibrate the trackin

Strany 11 - 1 Preparing for Use

3-48 Chapter3Functional TestingTracking Generator Level Flatness: HP E4401B and E4411B, Options 1DNand 1DQ7. Enter the appropriate power sensor Cal Fa

Strany 12 - Initial Inspection

Chapter 3 3-49Functional TestingTracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B,E4407B and E4408B, Option 1DNTracking Generator Leve

Strany 13

1-11 Preparing for UseThis chapter provides information on how to inspect your analyzerwhen it arrives and what to expect when it is first powered on.

Strany 14 - Power Requirements

3-50 Chapter3Functional TestingTracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B,E4407B and E4408B, Option 1DNFigure 3-8Procedure1. Ca

Strany 15 - AC Power Cable

Chapter 3 3-51Functional TestingTracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B,E4407B and E4408B, Option 1DN9. Repeat steps 5 throu

Strany 16 - Preparing for Use

4-14 Troubleshooting

Strany 17 - Preparation

4-2 Chapter4TroubleshootingWhat You Will Find in This ChapterWhat You Will Find in This ChapterThis chapter includes information on how to check for a

Strany 18

Chapter 4 4-3TroubleshootingBefore You Call Agilent TechnologiesBefore You Call Agilent TechnologiesCheck the BasicsA problem can often be resolved by

Strany 19

4-4 Chapter4TroubleshootingBefore You Call Agilent Technologies❏ Perform an Align Now, All. If the analyzer is an Agilent TechnologiesE4402B, E4403B,

Strany 20 - 2 Getting Started

Chapter 4 4-5TroubleshootingBefore You Call Agilent TechnologiesYou can order the service documentation, Option 0BW (assembly leveltroubleshooting and

Strany 21 - Getting Started

4-6 Chapter4TroubleshootingBefore You Call Agilent TechnologiesTable 4-1 Agilent Technologies Sales and Service OfficesUNITED STATESInstrument Support

Strany 22 - Front-Panel Features

Chapter 4 4-7TroubleshootingHow to Return Your Analyzer for ServiceHow to Return Your Analyzer for ServiceService TagIf you are returning your analyze

Strany 23

4-8 Chapter4TroubleshootingHow to Return Your Analyzer for ServiceFigure 4-1

Strany 24

1-2 Chapter1Preparing for UseInitial InspectionInitial InspectionInspect the shipping container and verify that the contents arecomplete. Keep the con

Strany 25

Chapter 4 4-9TroubleshootingHow to Return Your Analyzer for ServiceOther PackagingCAUTION Analyzer damage can result from using packaging materials ot

Strany 26

4-10 Chapter4TroubleshootingError MessagesError MessagesThe analyzer can generate various messages that appear on the displayduring operation. There a

Strany 27 - Rear-Panel Features

Chapter 4 4-11TroubleshootingStatus MessagesStatus MessagesThe following messages indicate a condition that may result inerroneous data being displaye

Strany 28 - Agilent E4408B

4-12 Chapter4TroubleshootingStatus MessagesE4411B only: disconnect any signals from the INPUTprior to performing this procedure.Align RF Skipped (Alig

Strany 29 - E4404B, E4405B and E4407B

Chapter 4 4-13TroubleshootingStatus MessagesInput is internal (no corresponding statusbit)This message applies to the Agilent TechnologiesE4401B and E

Strany 30

4-14 Chapter4TroubleshootingStatus MessagesOverload: Reduce Signal and press <ESC> (InputOverload Tripped)This message applies to the Agilent Te

Strany 31 - 2-12 Chapter2

Chapter 4 4-15TroubleshootingStatus MessagesSource Unlevel (Source Unleveled)Indicates the source power is set higher or lower thanthe analyzer can pr

Strany 32 - Display Annotation

4-16 Chapter4TroubleshootingInformational MessagesInformational MessagesThe following messages provide information that requires nointervention. The i

Strany 33 - Table 2-3 Screen Annotation

Chapter 4 4-17TroubleshootingInformational MessagesAtten auto set to 15 dBIndicates that an input signal has been detected whichis of sufficient level

Strany 34

4-18 Chapter4TroubleshootingInformational MessagesDefault spur table values loaded.No spur table has been previously saved when theOut-of-Band Spuriou

Strany 35

Chapter 1 1-3Preparing for UseInitial InspectionTable 1-1 Accessories Supplied with the AnalyzerDescriptionHP/Agilent PartNumberCommentsAdapter, Type-

Strany 36

Chapter 4 4-19TroubleshootingInformational MessagesMeasurement halted. Press a measurement key tocontinue.This error occurs after you choose Cancel to

Strany 37 - Key Overview

4-20 Chapter4TroubleshootingInformational MessagesTable saved successfully.This message appears after the user presses theSave Table key on the second

Strany 38

Chapter 4 4-21TroubleshootingError QueuesError QueuesWhen a user-error condition occurs in the instrument as a result ofSCPI (remote interface) activi

Strany 39 - Making a Measurement

4-22 Chapter4TroubleshootingError Message FormatError Message FormatThe system–defined error numbers are chosen on an enumerated (“1 ofN”) basis. The e

Strany 40

Chapter 4 4-23TroubleshootingError Message TypesError Message TypesEvents do not generate more than one type of error. For example, anevent that gener

Strany 41 - Measurement Summary

4-24 Chapter4TroubleshootingError Message Types201 to 799:Device-SpecificErrorsThese errors indicate that a device operation did not properly complete,

Strany 42 - Chapter 2 2-23

Chapter 4 4-25Troubleshooting0: No Error0:No Error0 No errorThe queue is empty. Every error in the queue has beenread or the queue was purposely clea

Strany 43 - Warm-up Time

4-26 Chapter4Troubleshooting–499 to –400: Query Errors–499 to –400:Query ErrorsThe instrument output queue control has detected a problem with themes

Strany 44 - File Menu Functions

Chapter 4 4-27Troubleshooting–499 to –400: Query Errors–440 Query UNTERMINATED after indefinite responseIndicates that a query was received in the sa

Strany 45 - 2-26 Chapter2

4-28 Chapter4Troubleshooting–399 to –300: Device-Specific Error Messages–399 to –300:Device-Specific Error MessagesAn error number in the range [−399 t

Strany 46 - Creating a directory

1-4 Chapter1Preparing for UsePower RequirementsPower RequirementsThe analyzer is a portable instrument and requires no physicalinstallation other than

Strany 47 - Format a Floppy Disk

Chapter 4 4-29Troubleshooting–299 to –200: Execution Error Messages–299 to –200:Execution Error MessagesAn error number in the range [–299 to –200] i

Strany 48 - Save a File

4-30 Chapter4Troubleshooting–199 to –100: Command Errors–199 to –100:Command ErrorsThe instrument parser detected an IEEE 488.2 syntax error. Errors

Strany 49

Chapter 4 4-31Troubleshooting–199 to –100: Command Errors–100 Command errorThis is a generic syntax error for devices that cannotdetect more specific

Strany 50 - Chapter 2 2-31

4-32 Chapter4Troubleshooting–199 to –100: Command Errors–114 Header suffix out of rangeThe value of a header suffix attached to a programmnemonic make

Strany 51

Chapter 4 4-33Troubleshooting–199 to –100: Command Errors–151 Invalid string dataA string data element was expected, but was invalid(see IEEE 488.2,

Strany 52 - Load a file

4-34 Chapter4Troubleshooting–199 to –100: Command Errors–150 String data errorThis error, as well as errors –151 through –159, isgenerated when parsi

Strany 53

Chapter 4 4-35Troubleshooting201 to 799: Device-Specific Errors201 to 799:Device-Specific ErrorsSome device operations did not properly complete, possi

Strany 54 - Rename a File

4-36 Chapter4Troubleshooting201 to 799: Device-Specific Errors219 Command not valid in this modelIndicates that the command sent from the remoteinterf

Strany 55 - Copy a File

Chapter 4 4-37Troubleshooting201 to 799: Device-Specific Errors607 File Name ErrorAn invalid file name has been specified. Use filenameswith a maximum of

Strany 56 - Delete a File

4-38 Chapter4Troubleshooting201 to 799: Device-Specific Errors729 In <filename>: error reading fileAppears when loading data from a limit line o

Strany 57 - Printer Setup and Operation

Chapter 1 1-5Preparing for UsePower RequirementsAC Power CableThe analyzer is equipped with a three-wire power cable, in accordancewith international

Strany 58

Chapter 4 4-39Troubleshooting201 to 799: Device-Specific Errors763 Incorrect filename, allowable extensions are.gif or .wmfAttempt to save a screen im

Strany 59 - Testing Printer Operation

4-40 Chapter4Troubleshooting201 to 799: Device-Specific Errors769 Invalid instrument modeYou have attempted to switch to an instrument modethat is cur

Strany 60 - Analyzer Battery Information

Chapter 4 4-41Troubleshooting201 to 799: Device-Specific Errors224 Option not licensed.The selected option requires a license. Refer to theinstallatio

Strany 61 - 3 Functional Testing

4-42 Chapter4Troubleshooting201 to 799: Device-Specific Errors223 Trigger Offset unavailable in swept spansTrigger Offset is only available in Zero Sp

Strany 62 - Functional Testing

Chapter 4 4-43Troubleshooting201 to 799: Device-Specific Errors752 Unable to load state from fileLoading of state from a file failed.755 Unable to load

Strany 63 - Test Descriptions

4-44 Chapter4Troubleshooting201 to 799: Device-Specific Errors207 Unable to store flatness dataA failure occurred in setting the flatness data in thein

Strany 64 - Displayed Average Noise Level

Chapter 4 4-45TroubleshootingGreater than 1000: Personality Specific Error MessagesGreater than 1000:Personality Specific Error MessagesAn error detect

Strany 65

4-46 Chapter4TroubleshootingGreater than 1000: Personality Specific Error Messages10168 Cannot update the list of cable types on driveC:The file update

Strany 66

Chapter 4 4-47TroubleshootingGreater than 1000: Personality Specific Error Messages3. Center the signal. (There must be a point on thetrace that is “X

Strany 67

4-48 Chapter4TroubleshootingGreater than 1000: Personality Specific Error Messages10251 Error reading file: RHOPMCO. Please reinstallcdmaOne.The file i

Strany 68

1-6 Chapter1Preparing for UsePower Requirements

Strany 69

Chapter 4 4-49TroubleshootingGreater than 1000: Personality Specific Error Messages10161 Lower Custom Mask is Invalid!The user-specified lower custom m

Strany 70

4-50 Chapter4TroubleshootingGreater than 1000: Personality Specific Error Messages10149 Opt BAH DSP algorithm code file requiresupgrade.The Digital Si

Strany 71

Chapter 4 4-51TroubleshootingGreater than 1000: Personality Specific Error Messages10239 Opt Freq Ref setting does not match externalreference.This me

Strany 72

4-52 Chapter4TroubleshootingGreater than 1000: Personality Specific Error Messages10172 Sweep Time too fast(<2sec)The sweep time must be set to 2 s

Strany 73

Chapter 4 4-53TroubleshootingGreater than 1000: Personality Specific Error Messages10170 The Cable Fault Measurement is active. ModeSetup is disabled.

Strany 74 - E4407B, E4408B

4-54 Chapter4TroubleshootingGreater than 1000: Personality Specific Error Messages10138 Valid GSM burst not found in frame (BurstType).No active GSM b

Strany 75

5-15 Menu MapsThis chapter provides a visual representation of the front panel keysand their associated menu keys. Refer to the Chapter 6 , “Front-Pan

Strany 76

5-2 Chapter5Menu MapsWhat You Will Find in This ChapterWhat You Will Find in This ChapterThis chapter provides menu maps for the front-panel keys havi

Strany 77

Chapter 5 5-3Menu MapsWhat You Will Find in This Chapter

Strany 78

5-4 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 79

Chapter 1 1-7Preparing for UsePreparationPreparationWARNING If this product is to be energized via an externalautotransformer for voltage reduction, m

Strany 80

Chapter 5 5-5Menu MapsWhat You Will Find in This Chapter

Strany 81

5-6 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 82 - Table 3-1

Chapter 5 5-7Menu MapsWhat You Will Find in This Chapter

Strany 83 - Frequency Readout Accuracy

5-8 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 84

Chapter 5 5-9Menu MapsWhat You Will Find in This Chapter

Strany 85 - Chapter 3 3-25

5-10 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 86 - Marker Count Accuracy

Chapter 5 5-11Menu MapsWhat You Will Find in This Chapter

Strany 87 - Frequency Response

5-12 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 88

Chapter 5 5-13Menu MapsWhat You Will Find in This Chapter

Strany 89

5-14 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 90

1-8 Chapter1Preparing for UseTurning On the Analyzer for the First TimeTurning On the Analyzer for the First TimeBefore using your analyzer, plug the

Strany 91

Chapter 5 5-15Menu MapsWhat You Will Find in This Chapter

Strany 92 - Reference Level Accuracy

5-16 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 93

Chapter 5 5-17Menu MapsWhat You Will Find in This Chapter

Strany 94 - Log Mode Procedure

5-18 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 95 - Linear Mode Procedure

Chapter 5 5-19Menu MapsWhat You Will Find in This Chapter

Strany 96

5-20 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 97

Chapter 5 5-21Menu MapsWhat You Will Find in This Chapter

Strany 98

5-22 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 99 - Table 3-6 Test Limits Table

Chapter 5 5-23Menu MapsWhat You Will Find in This Chapter

Strany 100 - Scale Fidelity

5-24 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 101 - Procedure

Chapter 1 1-9Preparing for UseTurning On the Analyzer for the First TimeNOTE Record the firmware revision number and keep it for reference. If youshoul

Strany 102

Chapter 5 5-25Menu MapsWhat You Will Find in This Chapter

Strany 103 - Test Description

5-26 Chapter5Menu MapsWhat You Will Find in This Chapter

Strany 104 - Required Equipment

Chapter 5 5-27Menu MapsWhat You Will Find in This Chapter

Strany 105

6-16 Front-Panel Key ReferenceThe front-panel keys in this chapter are listed alphabetically and aredescribed with their associated menu keys. The men

Strany 106 - Test Limits

6-2 Chapter6Front-Panel Key ReferenceKey LabelPageLocation< Display Line page 6-64> Display Line page 6-641 ↔ 2 page 6-1051 → 3 page 6-1052 → 3

Strany 107

Chapter 6 6-3Front-Panel Key ReferenceCorrections page 6-9Corrections page 6-33Corrections On Off page 6-9Counter Zoom On Off page 6-53Coupling AC DC

Strany 108

6-4 Chapter6Front-Panel Key ReferenceGate Delay page 6-85Gate Length page 6-85Gate On Off page 6-84Gate Setup page 6-84Graticule On Off page 6-24Harmo

Strany 109 - Option 1DN

Chapter 6 6-5Front-Panel Key ReferenceMkr → CF Step page 6-49Mkr → Ref Lvl page 6-49Mkr → Start page 6-49Mkr → Stop page 6-49Mkr ∆→ Span page 6-49MODE

Strany 110

6-6 Chapter6Front-Panel Key ReferenceRef Lvl Offst page 6-9Ref Level page 6-8Remote Port page 6-95Rename page 6-34Resolution Auto Man page 6-36Resolut

Strany 111

Chapter 6 6-7Front-Panel Key ReferenceTrace page 6-105Trace + State page 6-28Trace 1 2 3 page 6-105Tracking Peak page 6-77Trig page 6-100Trig Delay On

Strany 112 - Troubleshooting

iiThe information contained in this document is subject to changewithout notice.Agilent Technologies makes no warranty of any kind with regard to this

Strany 113 - 4-2 Chapter4

2-12 Getting Started

Strany 114 - Check the Basics

6-8 Chapter6Front-Panel Key ReferenceAMPLITUDE Y ScaleAMPLITUDE Y ScaleActivates the reference level function and accesses the amplitude menukeys. Amp

Strany 115 - Service Options

Chapter 6 6-9Front-Panel Key ReferenceAMPLITUDE Y ScalePresel Center Agilent E4404B, E4405B, E4407B and E4408B only. Adjusts thefrequency of the prese

Strany 116 - Chapter 4 4-5

6-10 Chapter6Front-Panel Key ReferenceAMPLITUDE Y ScaleSelect Selects which set of correction factors is to be modified.Once the selection has been mad

Strany 117 - 4-6 Chapter4

Chapter 6 6-11Front-Panel Key ReferenceAMPLITUDE Y ScalePointAllows you to create or edit an amplitude-correctionfactor data point. Up to 200 points m

Strany 118 - Original Packaging

6-12 Chapter6Front-Panel Key ReferenceAMPLITUDE Y ScaleDelete PointAllows you to delete the amplitude-correction data forthe currently selected point.

Strany 119 - Figure 4-1

Chapter 6 6-13Front-Panel Key ReferenceAuto CoupleAuto CoupleCouples the following functions: resolution bandwidth, videobandwidth, sweep coupling (SR

Strany 120 - Other Packaging

6-14 Chapter6Front-Panel Key ReferenceBW/AvgBW/AvgActivates the resolution bandwidth function and accesses the menukeys that control the bandwidth fun

Strany 121 - Error Messages

Chapter 6 6-15Front-Panel Key ReferenceBW/AvgAverage TypeVideo PowerAverage Type (Video) initiates a digital averaging routine that averageson a point

Strany 122 - Status Messages

6-16 Chapter6Front-Panel Key ReferenceBW/AvgEMI Res BW Accesses the EMI Res BW menu keys and allows you to choose betweenbandwidths of120 kHz, 9 kHz a

Strany 123

Chapter 6 6-17Front-Panel Key ReferenceDet/DemodDet/DemodAccesses the menu keys controlling detector functions, demodulationfunctions, the speaker, an

Strany 124

2-2 Chapter2Getting StartedWhat You Will Find in This ChapterWhat You Will Find in This ChapterThis chapter introduces the basic functions of the anal

Strany 125 - ESC to reset the

6-18 Chapter6Front-Panel Key ReferenceDet/DemodDemod ViewOn Off When Demod View (On) is pressed, the vertical scaling ofthe display is in frequency, a

Strany 126 - Chapter 4 4-15

Chapter 6 6-19Front-Panel Key ReferenceDisplayDisplayAccesses menu keys that allow you to control what is displayed on theanalyzer, including titles,

Strany 127 - Informational Messages

6-20 Chapter6Front-Panel Key ReferenceDisplayLimits Fixed RelAllows you to choose fixed or relative limit lines. Thefixed (Fixed) type uses the current

Strany 128 - Chapter 4 4-17

Chapter 6 6-21Front-Panel Key ReferenceDisplayModify Accesses menu keys that allow you to modify individuallimit lines. Key Access:Display, LimitsLimi

Strany 129

6-22 Chapter6Front-Panel Key ReferenceDisplayEditAccesses the Edit menu keys which include Point,Frequency, (or Time if X Axis Units (Time) has beense

Strany 130

Chapter 6 6-23Front-Panel Key ReferenceDisplayif a new point, to allow Frequency to be entered for thenew point. Key Access:Display, Limits, Modify, M

Strany 131 - Load Table key

6-24 Chapter6Front-Panel Key ReferenceDisplayTitle Accesses the following Title menu keys which allows you to change orclear a title on your display.

Strany 132 - Error Queues

Chapter 6 6-25Front-Panel Key ReferenceEnterEnterTerminates and enters into the analyzer a numerical value that hasbeen entered from the front panel u

Strany 133 - Error Message Format

6-26 Chapter6Front-Panel Key ReferenceEscEscDeactivates the active function and blanks the active function text fromthe display. No data can be accide

Strany 134 - Error Message Types

Chapter 6 6-27Front-Panel Key ReferenceFileFileAccesses the menu keys used to view, save, load, and manage data on afloppy disk or the internal analyze

Strany 135 - 4-24 Chapter4

Chapter 2 2-3Getting StartedFront-Panel FeaturesFront-Panel FeaturesFigure 2-1 Front-Panel Feature Overview1 Viewing Angle keys allow you to adjust th

Strany 136 - No Error

6-28 Chapter6Front-Panel Key ReferenceFileSaveAccesses menu keys that allow you to save analyzer setups, states,traces, limits, corrections and screen

Strany 137 - Query Errors

Chapter 6 6-29Front-Panel Key ReferenceFilecan only be saved individually. Refer to the File, Sourcekey description.ScreenDisplays all previously save

Strany 138 - –499 to –400: Query Errors

6-30 Chapter6Front-Panel Key ReferenceFileWhen Type is set to Corrections, Source accesses theAntenna, Cable, Other and User menu keys, which allowyou

Strany 139 - Device-Specific Error Messages

Chapter 6 6-31Front-Panel Key ReferenceFileTraceDisplays all trace files you may wish to load (TRC andCSV). Traces can be loaded individually or as a g

Strany 140 - Execution Error Messages

6-32 Chapter6Front-Panel Key ReferenceFileDir Up Allows you to move up one directory level. If at the toplevel,Dir Up moves to the drive level, displa

Strany 141 - Command Errors

Chapter 6 6-33Front-Panel Key ReferenceFileSort Sort accesses a menu of keys that allow you view thefiles you wish to delete, according to a selected fi

Strany 142 - –199 to –100: Command Errors

6-34 Chapter6Front-Panel Key ReferenceFileDir From To Allows you to select the source and destinationdirectories for your copy on one or more drives.D

Strany 143 - 4-32 Chapter4

Chapter 6 6-35Front-Panel Key ReferenceFileName Accesses the Alpha Editor and allows you to enter afilename. The external keyboard can also be used toe

Strany 144 - Chapter 4 4-33

6-36 Chapter6Front-Panel Key ReferenceFreq CountFreq CountActivates the Marker Normal function if there are no markers, and thenturns the marker count

Strany 145 - 4-34 Chapter4

Chapter 6 6-37Front-Panel Key ReferenceFREQUENCY ChannelFREQUENCY ChannelActivates the center frequency function, and accesses the menu offrequency fu

Strany 146 - Device-Specific Errors

2-4 Chapter2Getting StartedFront-Panel Features4 FREQUENCY Channel, SPAN X Scale, and AMPLITUDE YScale are the three large keys that activate the prim

Strany 147

6-38 Chapter6Front-Panel Key ReferenceFREQUENCY ChannelWhen a frequency offset is entered, its value appears on the bottom ofthe display (as opposed t

Strany 148 - Chapter 4 4-37

Chapter 6 6-39Front-Panel Key ReferenceHelpHelpAccesses a short description of any front panel or menu key. AfterpressingHelp, an explanation of the n

Strany 149 - DATA or MERGE

6-40 Chapter6Front-Panel Key ReferenceInput/OutputInput/OutputInput Z Corr 50Ω75ΩSets the input impedance for voltage-to-power conversions. Theimpedan

Strany 150 - Chapter 4 4-39

Chapter 6 6-41Front-Panel Key ReferenceInput/OutputExt Mix Band Accesses the Ext Mix Band key menus and allows youto select one of the pre-defined band

Strany 151 - 4-40 Chapter4

6-42 Chapter6Front-Panel Key ReferenceInput/OutputSince two measurements are taken for each displaycycle, the display update rate is reduced. Key Acce

Strany 152

Chapter 6 6-43Front-Panel Key ReferenceInput/OutputMixer Bias (On) activates and allows adjustment of aninternal bias source for use with external mix

Strany 153 - 4-42 Chapter4

6-44 Chapter6Front-Panel Key ReferenceMarkerMarkerAccesses the marker control keys which select the type and number ofmarkers and turns them on and of

Strany 154

Chapter 6 6-45Front-Panel Key ReferenceMarkerNOTE The delta marker function permits signal-to-noise measurementsprovided the signal is a single spectr

Strany 155 - 4-44 Chapter4

6-46 Chapter6Front-Panel Key ReferenceMarkerKey Access: Marker, More 1 of 2Readout Accesses the following menu keys that allow you to change the activ

Strany 156 - Greater than 1000:

Chapter 6 6-47Front-Panel Key ReferenceMarkerNOTE You can use Delta and Marker Noise to indicate noise level relative tosinusoidal signal (signal to n

Strany 157 - 4-46 Chapter4

Chapter 2 2-5Getting StartedFront-Panel Features10 The Data Control Keys, which include the step keys,knob, and numeric keypad, allow you to change th

Strany 158

6-48 Chapter6Front-Panel Key ReferenceMarkerIn log detector mode, with the result reported in dBm ina 1 Hz bandwidth, the final reported value will be:

Strany 159 - 4-48 Chapter4

Chapter 6 6-49Front-Panel Key ReferenceMarker →Marker →Accesses the following marker function menu keys:Mkr → CF Sets the center frequency of the anal

Strany 160 - Chapter 4 4-49

6-50 Chapter6Front-Panel Key ReferenceMeas ControlMeas ControlAccesses a menu of keys that allow you to pause or restart one of themeasurements availa

Strany 161 - 4-50 Chapter4

Chapter 6 6-51Front-Panel Key ReferenceMeas SetupMeas SetupIf in a measurement, displays one of five measurement setup menus.The setup menu displayed d

Strany 162 - Chapter 4 4-51

6-52 Chapter6Front-Panel Key ReferenceMeas SetupOccupied BW Meas Setup Menu KeysTo access the following keys for setting up an occupied bandwidthmeasu

Strany 163 - Save Table key to save a

Chapter 6 6-53Front-Panel Key ReferenceMeas SetupAvg NumberOn OffPress Avg Number (On) to specify the number of sweep averages usedwhen calculating th

Strany 164

6-54 Chapter6Front-Panel Key ReferenceMEASUREMEASUREIn the Spectrum Analyzer (SA) mode, accesses a menu of keys thatallow you to make adjacent channel

Strany 165 - 4-54 Chapter4

Chapter 6 6-55Front-Panel Key ReferenceMEASUREOccupied BW Integrates the power of the displayed spectrum and puts markers atthe frequencies between wh

Strany 166 - 5 Menu Maps

6-56 Chapter6Front-Panel Key ReferenceMEASUREWhen measuring the Nth harmonic, the analyzer will choose thenarrowest resolution bandwidth that is ≥N ti

Strany 167

Chapter 6 6-57Front-Panel Key ReferenceMODEMODESelects the measurement mode of your analyzer. Spectrum analyzermode (SA) is the default mode. Other mo

Strany 168 - Menu Maps

2-6 Chapter2Getting StartedFront-Panel FeaturesThe Step Keys (⇓ ⇑) increase or decrease the activefunction value. The step size depends upon the curre

Strany 169 - 5-4 Chapter5

6-58 Chapter6Front-Panel Key ReferenceMODE SETUPMODE SETUPBrings up a menu for setting up the current mode. The SA mode has nomode setup functions.

Strany 170 - Chapter 5 5-5

Chapter 6 6-59Front-Panel Key ReferenceNext WindowNext WindowAllows you to select the active window in functions which supportsplit-screen display mod

Strany 171 - 5-6 Chapter5

6-60 Chapter6Front-Panel Key ReferencePeak SearchPeak SearchPlaces a marker on the highest priority peak based upon one of twocriteria:Max (default) o

Strany 172 - Chapter 5 5-7

Chapter 6 6-61Front-Panel Key ReferencePeak SearchMarker NoiseReads out the average noise level, referenced to a 1 Hznoise power bandwidth. If the mar

Strany 173 - 5-8 Chapter5

6-62 Chapter6Front-Panel Key ReferencePeak SearchThe detection mode also affects the measurement. If inlog mode, the log detector understates the nois

Strany 174 - Chapter 5 5-9

Chapter 6 6-63Front-Panel Key ReferencePeak SearchNOTE For example, N dB Points can be used to measure the 3 dB bandwidthof a filter in a transmission

Strany 175 - 5-10 Chapter5

6-64 Chapter6Front-Panel Key ReferencePeak SearchThe value of the threshold appears in theactive-function block and on the lower-left side of thedispl

Strany 176 - Chapter 5 5-11

Chapter 6 6-65Front-Panel Key ReferencePresetPresetProvides a convenient starting point for making most measurements.You can select whether a factory

Strany 177 - 5-12 Chapter5

6-66 Chapter6Front-Panel Key ReferencePresetNOTE If the Power On function is set to Preset and the Preset function is set toFactory, turning on the an

Strany 178 - Chapter 5 5-13

Chapter 6 6-67Front-Panel Key ReferencePresetReference level 0 dBm in power-on unitsReference level offset 0 dBReference level position top (10th) gra

Strany 179 - 5-14 Chapter5

Chapter 2 2-7Getting StartedFront-Panel Features20 The Next Window key can be used to select the activewindow in functions which support split-screen

Strany 180 - Chapter 5 5-15

6-68 Chapter6Front-Panel Key ReferencePresetUser Preset Loads the analyzer configuration that existed when Save User Presetwas pressed. If Save User Pr

Strany 181 - 5-16 Chapter5

Chapter 6 6-69Front-Panel Key ReferencePrintPrintOption A4H (GPIB and Parallel) and 1AX (RS-232 and Parallel) only.Print initiates an output of the di

Strany 182 - Chapter 5 5-17

6-70 Chapter6Front-Panel Key ReferencePrint SetupPrint SetupOption A4H (GPIB and Parallel) or 1AX (RS-232 and Parallel) only.Accesses the menu keys wh

Strany 183 - 5-18 Chapter5

Chapter 6 6-71Front-Panel Key ReferencePrint SetupOrientation Allows you to select either Portrait or Landscape printing. TheOrientation key will not

Strany 184 - Chapter 5 5-19

6-72 Chapter6Front-Panel Key ReferenceRestartRestartRestarts the current measurement activated in the Measure key menu.When inAverage (On) mode, (BW/A

Strany 185 - 5-20 Chapter5

Chapter 6 6-73Front-Panel Key ReferenceReturnReturnReturns you to the previous menu. Repeated presses of this key moveback through previously selected

Strany 186 - Chapter 5 5-21

6-74 Chapter6Front-Panel Key ReferenceSaveSaveExecutes a save operation as though you were in the File, Save menuand had pressedSave Now. The Save har

Strany 187 - 5-22 Chapter5

Chapter 6 6-75Front-Panel Key ReferenceSingleSingleWhen analyzer is in continuous sweep mode and not in a measurement(Measure, Meas Off), this key cha

Strany 188 - Chapter 5 5-23

6-76 Chapter6Front-Panel Key ReferenceSourceSourceAccesses the tracking generator key functions (Options 1DN or 1DQonly). Without Option 1DN or 1DQ, p

Strany 189 - 5-24 Chapter5

Chapter 6 6-77Front-Panel Key ReferenceSourceAttenuationAuto ManAllows you to select between automatic and manual adjustment of thetracking generator’

Strany 190 - Chapter 5 5-25

2-8 Chapter2Getting StartedRear-Panel FeaturesRear-Panel FeaturesFigure 2-2 Rear-Panel Feature Overview1 Power input is the input for the AC line powe

Strany 191 - 5-26 Chapter5

6-78 Chapter6Front-Panel Key ReferenceSPAN X ScaleSPAN X ScaleActivates the Span function and accesses the menu of span functions.PressingSPAN X Scale

Strany 192 - Chapter 5 5-27

Chapter 6 6-79Front-Panel Key ReferenceSPAN X ScaleNOTE Zero Span is not available when Segmented (On) is selected. Segmentsin the segmented sweep fun

Strany 193 - Front-Panel Key Reference

6-80 Chapter6Front-Panel Key ReferenceSPAN X ScalePressing Zone (Off) will return you to a one-windowdisplay of the active window. Key Access:SPAN X S

Strany 194

Chapter 6 6-81Front-Panel Key ReferenceSPAN X ScaleNOTE Zone Span is not available when Segmented (On) is selected.Zone Pk Right Finds the next peak t

Strany 195

6-82 Chapter6Front-Panel Key ReferenceStandbyStandbyRemoves power from the analyzer, except for a small portion of circuitryinside the switching power

Strany 196

Chapter 6 6-83Front-Panel Key ReferenceSweepSweepActivates the sweep time function and accesses the following menukeys:Sweep TimeAuto ManSelects the l

Strany 197

6-84 Chapter6Front-Panel Key ReferenceSweepSweep ContSingleSwitches the analyzer between the continuous-sweep mode and thesingle-sweep mode. PressingS

Strany 198

Chapter 6 6-85Front-Panel Key ReferenceSweepEdge Setup Accesses menu keys that allow you to set up edgetriggering. Key Access:Sweep, Gate SetupEdge Po

Strany 199

6-86 Chapter6Front-Panel Key ReferenceSweepNOTE By selecting a number of sweep points greater than 401, you areoptimizing frequency resolution and acc

Strany 200 - AMPLITUDE Y Scale

Chapter 6 6-87Front-Panel Key ReferenceSweepSegmented Sweep EditorWhen the editor is entered (Sweep, Segmented, Modify, Edit), thesegmented sweep mode

Strany 201

Chapter 2 2-9Getting StartedRear-Panel Features4 Service Connector. The service connector is forservice use only.5 Inputs/Outputs (Refer to Agilent Te

Strany 202

6-88 Chapter6Front-Panel Key ReferenceSweepInteraction with Other Analyzer FunctionsCertain conditions prevent segmented sweep from being activated,wh

Strany 203

Chapter 6 6-89Front-Panel Key ReferenceSweepNOTE Turn markers off (Marker, Off) when entering or exiting segmentedsweep (Segmented, (On)(Off)).Perform

Strany 204

6-90 Chapter6Front-Panel Key ReferenceSweepAllows you to select a frequency range symmetricallyabout the center frequency for the current segmentusing

Strany 205 - Auto Couple

Chapter 6 6-91Front-Panel Key ReferenceSweepDelete SegmentDeletes the highlighted segment. Key Access: Sweep,Segmented, Modify, Edit, More 1 of 2Delet

Strany 206

6-92 Chapter6Front-Panel Key ReferenceSystem (Local)System (Local)Accesses the System menu keys. Pressing System (Local) after theanalyzer has been pl

Strany 207

Chapter 6 6-93Front-Panel Key ReferenceSystem (Local)Pressing Preset (User) sets the analyzer to the settingsdefined by theSave User Preset key. Refer

Strany 208

6-94 Chapter6Front-Panel Key ReferenceSystem (Local)Align Now Accesses the Align Now menu keys and immediatelyexecutes an alignment cycle of the selec

Strany 209 - Det/Demod

Chapter 6 6-95Front-Panel Key ReferenceSystem (Local)Remote Port With Option A4H (GPIB and Parallel) only. Selects the remote portto allow programming

Strany 210

6-96 Chapter6Front-Panel Key ReferenceSystem (Local)Optical Filter Accommodates the use of protective goggles whileviewing lasers. Key Access:System,

Strany 211

Chapter 6 6-97Front-Panel Key ReferenceSystem (Local)Licensing Accesses the security system to enable licensing for individual options.For information

Strany 212

2-10 Chapter2Getting StartedRear-Panel Features6 GPIB and parallel (Option A4H) is an optionalinterface. GPIB supports remote instrument operation.A p

Strany 213

6-98 Chapter6Front-Panel Key ReferenceSystem (Local)Activate Allows you to activate the option. When a valid optionkey number and license key number h

Strany 214

Chapter 6 6-99Front-Panel Key ReferenceSystem (Local)Install Allows you to install personalities in the analyzer. Formore information on this key, see

Strany 215

6-100 Chapter6Front-Panel Key ReferenceTrigTrigAccesses the menu of keys that allow you to select the sweep mode andtrigger mode of the analyzer.Free

Strany 216

Chapter 6 6-101Front-Panel Key ReferenceTrigTV Trig Setup Option B7B (TV Trigger and Picture on Screen) only. Accesses the setupfunctions for TV Trigg

Strany 217 - Chapter 6 6-25

6-102 Chapter6Front-Panel Key ReferenceTrigField OneCauses the selected line number to be viewed as anoffset into the first field starting with Line 1,

Strany 218

Chapter 6 6-103Front-Panel Key ReferenceTrigTrig Delay On Off Allows you to set and turn on or off a delay, during which the analyzerwill wait to begi

Strany 219

6-104 Chapter6Front-Panel Key ReferenceViewing AngleViewing AngleControls the optimum viewing angle of the display. The viewing anglekeys are located

Strany 220

Chapter 6 6-105Front-Panel Key ReferenceView/TraceView/TraceAccesses the trace keys that allow you to store and manipulate traceinformation. Each trac

Strany 221

6-106 Chapter6Front-Panel Key ReferenceView/Trace1↔2 Exchanges the contents of the trace 1 register with thetrace 2 register and puts trace 1 and 2 in

Strany 222

Chapter 6 6-107Front-Panel Key ReferenceView/TraceCAUTION Trace 1 should be in clear write mode prior to setting normalize to on.The normalize functio

Strany 223

iiiWARNING If this product is not used as specified, the protection providedby the equipment could be impaired. This product must be usedin a normal co

Strany 224

Chapter 2 2-11Getting StartedRear-Panel FeaturesAUX VIDEO OUT provides detected video output (beforethe analog-to-digital conversion) proportional tov

Strany 225

6-108 Chapter6Front-Panel Key ReferenceZoomZoomAllows you to switch between the split-screen and full-sized display ofthe active window in zone span a

Strany 226

7-17 Options and AccessoriesThis chapter lists the options and accessories that can be ordered alongwith your analyzer. The Option Table on the follow

Strany 227

7-2 Chapter7Options and AccessoriesTable 7-1 Option TableOptionE4411BE4401BE4402BE4403BE4404BE4405BE4407BE4408B0B0 Delete Manual Set ••••••••0B1 Add M

Strany 228 - Freq Count

Chapter 7 7-3Options and AccessoriesB7D Digital Signal Processing andFast ADC••••B7E RF Communications Hardware ••••B7K Distance to Fault Accessory Ki

Strany 229 - FREQUENCY Channel

7-4 Chapter7Options and AccessoriesOptionsOptionsOptions can be ordered by option number when you order youranalyzer. Some of the options are also ava

Strany 230

Chapter 7 7-5Options and AccessoriesHow to Order OptionsHow to Order OptionsOptions can be ordered using your product number and the number ofthe opti

Strany 231

7-6 Chapter7Options and AccessoriesHow to Order OptionsRack Mount Kit With Handles (Option 1CP)Option 1CP provides the parts necessary to mount the an

Strany 232 - Input/Output

Chapter 7 7-7Options and AccessoriesHow to Order OptionsNarrow Resolution Bandwidth (Option 1DR)This option provides additional narrow resolution band

Strany 233

7-8 Chapter7Options and AccessoriesHow to Order OptionsAUX VIDEO OUT - provides detected video output (before theanalog-to-digital conversion) proport

Strany 234

Chapter 7 7-9Options and AccessoriesHow to Order OptionsAUX VIDEO OUT - provides detected video output (before theanalog-to-digital conversion) propor

Strany 235

2-12 Chapter2Getting StartedRear-Panel Features15 DC Fuse protects the analyzer from drawing too muchDC power. Replace only with a fuse of the same ra

Strany 236

7-10 Chapter7Options and AccessoriesHow to Order OptionsTV Trigger and Picture on Screen (Option B7B)Option B7B requires Option BAA.Option B7B allows

Strany 237

Chapter 7 7-11Options and AccessoriesHow to Order OptionsAPC 3.5 Input Connector (Option BAB)The type-N female connector is replaced with an APC 3.5 m

Strany 238

7-12 Chapter7Options and AccessoriesHow to Order OptionsFront Panel Protective Cover (Option UK9)The cover assembly snaps onto the front of your analy

Strany 239

Chapter 7 7-13Options and AccessoriesAccessoriesAccessoriesA number of accessories are available from Agilent Technologies to helpyou configure your an

Strany 240

7-14 Chapter7Options and AccessoriesAccessoriesHP/Agilent E1779A Battery PackThe E1779A is a battery pack that will power an Agilent ESA spectrumanaly

Strany 241 - Marker →

Chapter 7 7-15Options and AccessoriesAccessoriesExternal KeyboardThe HP/Agilent C1405B keyboard is an IBM AT compatible keyboardthat can be connected

Strany 242 - Meas Control

7-16 Chapter7Options and AccessoriesAccessoriesHP/Agilent 11970 Series Harmonic MixersThe 11970 Series harmonic mixers are available to extend thefreq

Strany 243 - Meas Setup

IndexIndex 1Symbols, 6-64(Local) key, 6-92> Display Line, 6-64Numerics10 MHz REF INPUT, 2-1110 MHz REF OUTPUT, 2-1150 ohm RF bridge, 7-1350 ohm/75

Strany 244

2 IndexIndexnegative peak, 6-17positive peak, 6-17sample, 6-17Detector, 6-17detector functions, 6-17device-specific error messages201 to 702, 4-35Diagn

Strany 245

IndexIndex 3Input key, 6-40Input menu map, 5-3Input Mixer Int Ext key, 6-40Input Mixer key, 6-40Input Z Corr, 6-40Input/Output key, 6-40instrument pre

Strany 246

Chapter 2 2-13Getting StartedDisplay AnnotationDisplay AnnotationHere is an example of the annotation that may appear on an analyzerdisplay. The displ

Strany 247

4 IndexIndexoverview, keys and key menus,2-18Ppackaging, 4-7parallel interface connector, 2-10Pause, 6-50peaknext, 6-62peak detection modepositive/neg

Strany 248

IndexIndex 5last, 6-79zero, 6-78span functions, 6-78Span key, 2-4Span Pair Span Center, 6-45SPAN X Scale, 6-78SPAN X Scale menu map, 5-3Span Zoom, 6-7

Strany 249 - Chapter 6 6-57

2-14 Chapter2Getting StartedDisplay AnnotationTable 2-3 Screen AnnotationItem Description Function Key1 Detector mode Detector2 Reference level Ref Le

Strany 250 - MODE SETUP

Chapter 2 2-15Getting StartedDisplay AnnotationWhen the analyzer is set to the external mixer state (Option AYZ), item6 changes to display Ext Mix in

Strany 251 - Next Window

2-16 Chapter2Getting StartedDisplay AnnotationItem 24 refers to the trace modes of the analyzer. The first letter Windicates that the analyzer is in cl

Strany 252 - Peak Search

Chapter 2 2-17Getting StartedDisplay AnnotationTable 2-7 Screen Annotation for GPIB AnnunciatorsScreenAnnotationDescriptionR Remote OperationL GPIB Li

Strany 253

2-18 Chapter2Getting StartedKey OverviewKey OverviewThe keys labeled FREQUENCY Channel, System, and Marker are allexamples of front-panel keys. Pressi

Strany 254

Chapter 2 2-19Getting StartedKey OverviewIn some menus of keys, one key label will always be highlighted to showwhich key has been selected but the me

Strany 255

2-20 Chapter2Getting StartedMaking a MeasurementMaking a MeasurementLet’s begin using the analyzer by measuring an input signal. Since theinternal 50

Strany 256

ivEXCLUSIVE REMEDIESTHE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE ANDEXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOTBE LIABLE FOR ANY DIRECT, INDIR

Strany 257

Chapter 2 2-21Getting StartedMaking a Measurement6. Set the amplitude.When the peak of a signal does not appear on the screen, it may benecessary to a

Strany 258

2-22 Chapter2Getting StartedMaking a Measurement7. Set the marker.Marker functions measure the frequency and amplitude of a signal.You can place a dia

Strany 259

Chapter 2 2-23Getting StartedMaking a MeasurementFigure 2-5 Reading the Amplitude and Frequency

Strany 260

2-24 Chapter2Getting StartedAnalyzer Accuracy and the Internal Alignment ProcessAnalyzer Accuracy and the Internal AlignmentProcessData from the inter

Strany 261

Chapter 2 2-25Getting StartedFile Menu FunctionsFile Menu FunctionsThis section describes how to use the functions located under thefront-panelFile ke

Strany 262 - Print Setup

2-26 Chapter2Getting StartedFile Menu FunctionsFigure 2-6 Catalog Menu1. The entry fields show the parameters for the files viewed. Entryfields that are

Strany 263

Chapter 2 2-27Getting StartedFile Menu Functionsd. Modified: column displays the date and time the last changeoccurred.3. The softkey menu has four op

Strany 264 - Measure key menu

2-28 Chapter2Getting StartedFile Menu FunctionsFormat a Floppy DiskYou can format a floppy disk in the analyzer. The format is MS-DOS1.Note that it is

Strany 265

Chapter 2 2-29Getting StartedFile Menu FunctionsFigure 2-7 Format MenuYou will see the message: Formatting Disk in the display window.Pressing any oth

Strany 266

2-30 Chapter2Getting StartedFile Menu Functions2. Set the center frequency to 50 MHz and the span to 75 MHz: pressFREQUENCY, 50 MHz, SPAN, 75 MHz.The

Strany 267

Contentsv1. Preparing for UseInitial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 268

Chapter 2 2-31Getting StartedFile Menu FunctionsFigure 2-9 Viewing Traces 1 and 25. Activate trace 3 by pressing Trace (3), Clear Write. Your analyzer

Strany 269

2-32 Chapter2Getting StartedFile Menu FunctionsStep 2. Save the file1. To access theSave menu, press, File, Save.2. Select the type of file you want to

Strany 270 - SPAN X Scale

Chapter 2 2-33Getting StartedFile Menu FunctionsLoad a file1. Reset the analyzer by pressing Preset, Factory Preset (if present).2. To access theLoad m

Strany 271

2-34 Chapter2Getting StartedFile Menu FunctionsNOTE You should notice that the trace you saved, in this example trace 3, isloaded to trace 2. If you w

Strany 272

Chapter 2 2-35Getting StartedFile Menu FunctionsRename a File1. To access the Rename menu, press File, Rename.2. Select the type of file you want to re

Strany 273

2-36 Chapter2Getting StartedFile Menu FunctionsCopy a File1. To access the Copy menu, press File, Copy. This menu functionbrings up 2 directory boxes

Strany 274 - 6-82 Chapter6

Chapter 2 2-37Getting StartedFile Menu FunctionsNOTE Highlighting a filename in the To: directory box has no significance.The file cannot be placed insid

Strany 275

2-38 Chapter2Getting StartedPrinter Setup and OperationPrinter Setup and OperationEquipment• Agilent ESA Spectrum Analyzer equipped with Options A4H (

Strany 276

Chapter 2 2-39Getting StartedPrinter Setup and Operation6. Press Printer Type to access the Printer Type menu keys. Press Auto tomake the analyzer att

Strany 277

2-40 Chapter2Getting StartedPrinter Setup and OperationTesting Printer OperationWhen you have completed the analyzer’s printer setup, press PrintSetup

Strany 278

ContentsviProcedure (10 MHz to 500 MHz) E4401B and E4411B . . . . . . . . . . . . . . . . . . . . . . . . 3-6Procedure (501 MHz to 1.0 GHz) E4401B an

Strany 279

Chapter 2 2-41Getting StartedAnalyzer Battery InformationAnalyzer Battery InformationThe analyzer uses a lithium battery to enable the internal memory

Strany 280 - 6-88 Chapter6

3-13 Functional Testing

Strany 281

3-2 Chapter3Functional TestingWhat You Will Find in This ChapterWhat You Will Find in This ChapterThis chapter describes the functional tests and prov

Strany 282

Chapter 3 3-3Functional TestingTest DescriptionsTest DescriptionsEach of the following test descriptions include the test limits (pass/failcriteria),

Strany 283

3-4 Chapter3Functional TestingDisplayed Average Noise LevelDisplayed Average Noise LevelTest LimitsFrequency Range Model (50 Ω Inputs) Maximum(50 Ω In

Strany 284 - System (Local)

Chapter 3 3-5Functional TestingDisplayed Average Noise LevelTest DescriptionThe Displayed Average Noise Level is measured within the frequencyrange sp

Strany 285

3-6 Chapter3Functional TestingDisplayed Average Noise LevelProcedure (10 MHz to 500 MHz) E4401B and E4411B1. Connect the equipment as shown in Figure

Strany 286

Chapter 3 3-7Functional TestingDisplayed Average Noise Level7. Press the following keys on the analyzer:BW/Avg, 1 kHzVideo BW Auto Man, 30 HzSingleWai

Strany 287

3-8 Chapter3Functional TestingDisplayed Average Noise Level4. Press the following keys on the analyzer:Sweep, Sweep (Cont)BW/Avg, Average (Off)BW/Avg,

Strany 288

Chapter 3 3-9Functional TestingDisplayed Average Noise LevelProcedure (1.01 GHz to 1.5 GHz) E4401B and E4411B1. Press the following keys on the analyz

Strany 289

ContentsviiProcedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-41Second

Strany 290

3-10 Chapter3Functional TestingDisplayed Average Noise Level7. Press the following keys on the analyzer:Display, Display Line (On)Adjust the display

Strany 291

Chapter 3 3-11Functional TestingDisplayed Average Noise Level5. Press Center Freq. Set the center frequency of the analyzer to thefrequency recorded i

Strany 292

3-12 Chapter3Functional TestingDisplayed Average Noise Level4. Press the following keys on the analyzer:Sweep, Sweep (Cont)BW/Avg, Average (Off)BW/Avg

Strany 293

Chapter 3 3-13Functional TestingDisplayed Average Noise Level2. Press the following keys on the analyzer:SingleBW/Avg, Average (On), 3, EnterSingleWai

Strany 294

3-14 Chapter3Functional TestingDisplayed Average Noise LevelProcedure (3.01 GHz to 6.0 GHz) E4404B, E4405B,E4407B, E4408B1. Press System, Power On/Pre

Strany 295

Chapter 3 3-15Functional TestingDisplayed Average Noise Level7. Press the following keys on the analyzer:Display, Display Line (On)Adjust the display

Strany 296 - Viewing Angle

3-16 Chapter3Functional TestingDisplayed Average Noise Level5. Press Center Freq. Set the center frequency of the analyzer to thefrequency recorded in

Strany 297 - View/Trace

Chapter 3 3-17Functional TestingDisplayed Average Noise Level4. Press the following keys on the analyzer:Sweep, Sweep (Cont)BW/Avg, Average (Off)BW/Av

Strany 298

3-18 Chapter3Functional TestingDisplayed Average Noise Level2. Press the following keys on the analyzer:SingleBW/Avg, Average (On), 3, EnterSingleWait

Strany 299

Chapter 3 3-19Functional TestingDisplayed Average Noise LevelProcedure (12.01 GHz to 22 GHz) E4407B and E4408B1. Press System, Power On/Preset, Preset

Strany 300

viiiContents–299 to –200:Execution Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-

Strany 301 - 7 Options and Accessories

3-20 Chapter3Functional TestingDisplayed Average Noise Level7. Press the following keys on the analyzer:Display, Display Line (On)Adjust the display

Strany 302 - Table 7-1 Option Table

Chapter 3 3-21Functional TestingDisplayed Average Noise Level4. Press the following keys on the analyzer:Sweep, Sweep (Cont)BW/Avg, Average (Off)BW/Av

Strany 303

3-22 Chapter3Functional TestingDisplayed Average Noise LevelTable 3-1Model Number Frequency RangeMeasuredFrequencyAverageNoise Level(TR Entry)Maximum5

Strany 304 - Options and Accessories

Chapter 3 3-23Functional TestingFrequency Readout AccuracyFrequency Readout AccuracyTest LimitsTest DescriptionThe frequency readout accuracy of the a

Strany 305 - How to Order Options

3-24 Chapter3Functional TestingFrequency Readout AccuracyFigure 3-2Procedure1. Connect the equipment as shown in Figure 3-2. Remember toconnect the 10

Strany 306 - 7-6 Chapter7

Chapter 3 3-25Functional TestingFrequency Readout Accuracy5. Repeat step 3. If the instrument is functioning correctly, the resultshould be between 1.

Strany 307 - Pre-amplifier (Option 1DS)

3-26 Chapter3Functional TestingMarker Count AccuracyMarker Count AccuracyProcedure1. Press System, Power On/Preset, Preset (Factory), Preset on theana

Strany 308 - 7-8 Chapter7

Chapter 3 3-27Functional TestingFrequency ResponseFrequency ResponseTest LimitsTest DescriptionThe output of the synthesized sweeper is fed through a

Strany 309 - Chapter 7 7-9

3-28 Chapter3Functional TestingFrequency ResponseRequired EquipmentInstrumentCriticalSpecifications(for this test)RecommendedHP/Agilent ModelSignal Sou

Strany 310 - FM Demodulation (Option BAA)

Chapter 3 3-29Functional TestingFrequency ResponseFigure 3-3Procedure1. Zero and calibrate the power meter and power sensor as described inthe power m

Strany 311 - Chapter 7 7-11

ContentsixSource . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-76S

Strany 312 - 7-12 Chapter7

3-30 Chapter3Functional TestingFrequency ResponseVideo BW Auto Man, 3 kHzPeak SearchFREQUENCY, Signal Track (On)5. Adjust the synthesized sweeper ampl

Strany 313 - Accessories

Chapter 3 3-31Functional TestingFrequency ResponseTable 3-3 Frequency Response ReportModel CenterFreq AnalyzerAmplitudePowerMeterAmplitudeFlatnessErro

Strany 314 - AC Probe

3-32 Chapter3Functional TestingReference Level AccuracyReference Level AccuracyTest LimitsTest DescriptionA 50 MHz CW signal is applied to the Input o

Strany 315 - Chapter 7 7-15

Chapter 3 3-33Functional TestingReference Level AccuracyRequired EquipmentFigure 3-4Instrument Critical Specifications(for this test)RecommendedHP/Agil

Strany 316

3-34 Chapter3Functional TestingReference Level AccuracyLog Mode Procedure1. Connect the equipment as shown in Figure 3-4.2. PressPRESET on the synthes

Strany 317

Chapter 3 3-35Functional TestingReference Level AccuracyLinear Mode Procedure1. Set the power level on the synthesized sweeper to –30 dBm.2. Set the a

Strany 318

3-36 Chapter3Functional TestingReference Level Accuracy4. Now that the reference has been established in step 4, adjust thesynthesized sweeper power l

Strany 319

Chapter 3 3-37Functional TestingResolution Bandwidth Switching UncertaintyResolution Bandwidth Switching UncertaintyTest LimitsTest DescriptionTo meas

Strany 320

3-38 Chapter3Functional TestingResolution Bandwidth Switching UncertaintyRequired Equipment(No Equipment Required)ProcedureNOTE The 50 MHz reference o

Strany 321

Chapter 3 3-39Functional TestingResolution Bandwidth Switching Uncertainty5. Repeat step 3 and 4 for each of the remaining resolution bandwidthand spa