Agilent-technologies E8247C PSG CW Uživatelský manuál Strana 206

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 220
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 205
196 Chapter 10
Troubleshooting
Data Storage Problems
Data Storage Problems
Registers With Previously Stored Instrument States are Empty
The save/recall registers are backed-up by a battery when line power to the signal generator is not
connected. The battery may need to be replaced.
To verify that the battery has failed:
1. Turn off line power to the signal generator.
2. Unplug the signal generator from line power.
3. Plug in the signal generator.
4. Turn on the signal generator.
5. Observe the display for error messages.
If either error message 311 or 700 is stored in the error message queue, the signal generators battery
has failed.
6. Refer to the Service Guide for battery replacement instructions.
Saved Instrument State, but Register is Empty or Contains Wrong State
If you select a register number greater than 99, the signal generator automatically selects register 99 to save
the instrument state.
If the register number you intended to use is empty or contains the wrong instrument state, recall register 99:
Press
Recall > 99 > Enter.
The lost instrument state may be saved there.
Zobrazit stránku 205
1 2 ... 201 202 203 204 205 206 207 208 209 210 211 ... 219 220

Komentáře k této Příručce

Žádné komentáře

Case-mate Non manuels d'utilisation

La liste de Case-mate Non produits contient 1 manuels d'utilisation pour les modèles 1. Tous les guides d'utilisation en format pdf peuvent être téléchargés ou consultés en ligne.

Cliquez sur un alphabet ci-dessous pour faire défiler la liste complète des modèles commençant par cette lettre

Modèles Type de document
CM010092
Fiche technique   Case-mate iPhone 3G/3GS Battery Case, 6 pages