Features and Functions 2
U2761A User’s Guide 35
Output Units
This configuration applies to output amplitude only. At
power-on, the units for output amplitude are volts
peak-to-peak.
The output units consist of Vpp, Vrms, or dBm. The
default unit is Vpp.
The unit setting is stored in volatile memory. The units
are set to “Vpp” upon power-off or after a remote
interface reset.
The output units for amplitude cannot be set to dBm if
the output termination is currently set to
“high impedance”. The units are automatically converted
to Vpp.
Soft Front Panel Operation
On the Amplitude panel as shown in Figure 2- 4, input the
desired amplitude value and select the unit from the drop
down list.
Remote Interface Operation
VOLTage:UNIT {VPP|VRMS|DBM}