7-16 85330A Multiple Channel Controller
Remote Programming Command Reference
Standard Command Reference
RUNTime:TIMEout:REMote
Allows a timeout to occur if the REMote1 or REMote2 controller does not
respond in determined time.
Default The default is 1 second (1000000 microseconds).
Syntax RUNT:TIME:REM <microseconds>;
Example RUNT:TIME:REM 1000000;
NOTE This command is in revision A.02.31 or greater.
RUNTime:TIMEout:SOURce
Allows a timeout to occur if the source 1 or source 2 ready line does not
respond in determined time.
Default The default is 1 second (1000000 microseconds).
Syntax RUNT:TIME:SOUR <microseconds>;
Example RUNT:TIME:SOUR 1000000;
SYSTem
Subsystem Command Syntax
SYSTem
:ERRor?