Agilent Technologies 3000 X-Series Servisní příručka Strana 79

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 156
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 78
Capturing Data 3
Agilent 3000 Series Oscilloscopes User’s Guide 79
4 Press Slope to select the edge to trigger on:
— Rising edge.
— Falling edge.
To set up pulse width triggers
A pulse width trigger occurs when a pulse is found in a waveform that
matches the pulse definition.
1 Press Mode/Coupling
.
2 In the Trigger menu, press Mode until “Pulse” is selected.
3 Press Source to select the waveform to trigger on:
CH1
CH2
EXT — the external trigger input.
EXT/5 — the (5:1) attenuated external trigger input.
4 Press When to select the type of pulse to trigger on:
— Positive pulse greater than the width setting.
— Positive pulse less than the width setting.
— Positive pulse equal to the width setting.
— Negative pulse greater than the width setting.
— Negative pulse less than the width setting.
— Negative pulse equal to the width setting.
5 Press Setting and turn the Entry
knob to adjust the width setting.
The width setting can be adjusted from 20 ns to 10 s.
Zobrazit stránku 78