Agilent Technologies 4000 X Servisní příručka Strana 65

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 168
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 64
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs 3
4000 X-Series Oscilloscopes Advanced Training Guide 65
26 Press the [Trigger] front panel key.
27 Press the “runt polarity” softkey and select the “either polarity” (third)
icon.
28 Tap the Qualifier softkey to select None.
29 Press the [Acquire] front panel key.
30 Tap the Segmented softkey; then double-tap # of Segs to enter 500 as the
number of segments to capture.
31 Tap the Segmented softkey to begin a Segmented Memory acquisition.
32 Tap the Current Segment softkey; then turn the Entry knob to review all
500 captured segments.
33 Set the Current Segment to 500 and note the time-tag of the last captured
segment.
Segmented Memory optimizes oscilloscope acquisition memory by only
capturing important segments of a waveform based on the trigger
condition and timebase setting. In this example, we have selectively
captured 500 occurrences of positive and negative runts of any width for a
total acquisition time of more than 20 ms as shown in Figure 40.
Figure 40 Using the scope’s Segmented Memory acquisition to selectively capture 500
consecutive runt pulses.
Zobrazit stránku 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 167 168

Komentáře k této Příručce

Žádné komentáře