530 Chapter7
Language Reference
MKPT Marker Threshold
MKPT Marker Threshold
Syntax
Figure 7-177 MKPT Syntax
Description
The MKPT command sets the minimum amplitude level from which a
peak on the trace can be detected. The default value is 130 dBm. See
also the MKPX command.
Any portion of a peak that falls below the peak threshold is used to
satisfy the peak excursion criteria. For example, a peak that is equal to
3 dB above the threshold when the peak excursion is equal to 6 dB will
be found if the peak extends an additional 3 dB or more below the
threshold level.
Parameters
number real from 200 to 30.
UP or DN increments of 10 dB.
Query Response
Figure 7-178 MKPT Query Response