Source supplemental characteristics
Max output power and source/sink limits:
31.8 W
±6 V @ ±3.03 A, ±21 V @ ±1.515 A, ±210 V @ ±105 mA, four quadrant source or sink
operation.
Current limit/compliance:
Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA in
10 nA range.
Voltage limit/compliance:
Accuracy is same as voltage source. Minimum value is 1 % of range, or 20 mV in
200 mV range
Over range :
101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other than
1.5 A and 3 A ranges. No over range for 200 V range with current exceeding 105 mA
pulse only condition.
Over temperature protection:
Output turns off then resets at over temperature sensed internally.
Voltage output settling time:
Step is 10 % to 90 % range
Slew rate:
Current output settling time:
condition. Step is 10 % to 90 % range
10 nA, 100 nA ranges: < 10 ms
Noise 10 Hz to 20 MHz (V source): 3 mVrms, 20 V range.
V source overshoot:
< ±(0.1 % + 10 mV). Step is 10 % to 90 % range, resistive load.
I source overshoot:
< ±0.1 % (< ±0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load.
Voltage source range change overshoot:
Current source range change overshoot:
06 | Keysight | B2900A Series Precision Source/Measure Unit – Data Sheet