2- 32 Agilent B1500A User’s Guide, Edition 13
Introduction
Measurement Modules
MFCMU - Multi Frequency CMU
This section describes typical specification of the multi frequency capacitance
measurement unit (CMU). The CMU performs the impedance measurement and
returns the specified measurement data such as Cp-G. Only one module can be
installed in one mainframe.
Measurement parameters: See Table 2-15.
AC signal, output frequency:
1 kHz to 5 MHz
Setting resolution: 1 mHz (1 kHz to), 10 mHz (10 kHz to), 0.1 Hz (100 kHz to),
or 1 Hz (1 MHz to 5 MHz)
AC signal, output level:
10 mVrms to 250 mVrms, 1 mV resolution
•DC bias:
0 to ± 25 V, using MFCMU internal DC bias
0 to ± 100 V, using SMU and SCUU (SMU CMU unify unit)
Measurement range:
For the fixed ranging mode, measurement range (impedance range) must be
specified to perform measurement.
Table 2-16 shows the available measurement
ranges and the corresponding impedance value. And Figure 2-7 shows the
calculation example of the impedance vs frequency characteristics of a
capacitive load. Use these information to decide the measurement range.
Impedance Z is calculated by the following formula.
Z = 1 / (2 f C)
where f is frequency (Hz) and C is capacitance (F).
Zobrazit stránku 73