4-82
Screens
Output RF Spectrum, Main View (Option 006 Only)
Output RF Spectrum, Main View (Option 006 Only)
The Agilent 8922M/S uses a 3-pole synchronously tuned filter to make Output RF
Spectrum measurements rather than a 5-pole filter as specified in the GSM
recommendations. Refer to Section titled “Output RF Spectrum Measurements
Using a 3-Pole Synchronously Tuned Measurement Filter” in Chapter 2.
1. Amplitude This is a copy of the RF Analyzer Amplitude field. This is the amplitude to be
assumed at the selected input port, see RF Analyzer Functions).
Range RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0
dBm.
See Also Screens: RF Generator/RF Analyzer, (RF Analyzer)
2. Freq Offset This sets the frequency offset for the Output RF Spectrum measurement. This field
is used when not making reference measurements. The offset used is 0.0 kHz when
Mode is set to Ramp Ref or Mod Ref.
Range 2.0 MHz to +2.0 MHz.
3. Sync Status This field displays any errors that occurred while trying to synchronize to the
demodulated data.
See Also Making Measurements: Solving Problems Messages