Agilent 75000 Series B
Agilent E1346A
48-Channel Relay Multiplexer
Service Manual
Enclosed is the Service Manual for the Agilent E1346A
48-Channel Relay Multiplexer. Insert this manual, along
with any other VXIbus manuals that you have, into the binder
that came with your Agilent Technologies mainframe.
*E1346-90011*
Manual Part Number: E1346-90011
Printed in Malaysia E0706
Zobrazit stránku 0
1 2 3 4 5 6 ... 65 66

Shrnutí obsahu

Strany 1 - *E1346-90011*

Agilent 75000 Series BAgilent E1346A48-Channel Relay MultiplexerService Manual Enclosed is the Service Manual for the Agilent E1346A

Strany 2

SafetyConsiderationsThis product is a Safety Class I instrument that is provided with a protectiveearth terminal when installed in the mainframe. Che

Strany 3

WARNINGUSING AUTOTRANSFORMERS. If the mainframe is to beenergized via an autotransformer (for voltage reduction) makesure the common terminal is conn

Strany 4 - Safety Symbols

Relay Life Electromagnetic relays are subject to normal wear-out. Relay life dependson several factors. Two factors are loading and switching frequenc

Strany 5 - Manual Descriptions

ReplacementStrategyThe replacement strategy also depends on the application. If some relays areused more often, or at higher load, than the others, th

Strany 6 - What’s in this Manual

Multiplexer DescriptionThe Agilent E1346A Single Ended Relay Multiplexer is an "instrument" ina VXIbus mainframe. As such, the multiplexer i

Strany 7 - Contents

Multiplexer Serial NumbersMultiplexers covered by this manual are identified by a serial number prefixlisted on the title page. Agilent Technologies u

Strany 8

RecommendedTest EquipmentTable 1-1 lists the test equipment recommended for testing, adjusting, andservicing the relay multiplexer. Essential requirem

Strany 9 - General Information

Notify Agilent and carrier.Notify AgilentReturn Multiplexer to AgilentFigure 1-2. Initial (Incoming) Inspection Guidelines1-9 General Information

Strany 10 - Considerations

ShippingGuidelinesFollow the steps in Figure 1-3 to return a relay multiplexer to a AgilentTechnologies Sales and Support Office or Service Center.

Strany 11 - 1-3 General Information

2Verification TestsIntroduction This chapter describes the verification tests for the Agilent E1346A SingleEnded Relay Multiplexer. The three levels o

Strany 13 - Replacement

FunctionalVerification TestThe Functional Verification Test for the Agilent E1346A Single EndedRelay Multiplexer consists of sending the *IDN? command

Strany 14 - Description

PerformanceVerification TestsThe procedures in this section are used to test the multiplexer’s electricalperformance using the specifications in Appen

Strany 15 - 1-7 General Information

Test 2-1: ClosedChannelResistance TestThis test verifies that all relay contacts meet the closed-channel resistancespecification for the multiplexer.

Strany 16 - Shipping

Table 2-1. Measured Protection Resistor ValuesResistorMeasured Value (Ω)ResistorMeasured Value (Ω)R38R39R40___________________________________________

Strany 17 - 1-9 General Information

Ch 90-92LO Measurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM leads as shown in Figure 2-4• Turn mainframe power ON 1.

Strany 18 - Guidelines

Ch 90-92GU Measurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM leads as shown in Figure 2-5• Turn mainframe power ON 2.

Strany 19 - Verification Tests

Ch 93 HI Measurement 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM leads as shown in Figure 2-6• Turn mainframe power ON2. Mea

Strany 20 - 2-2 Verification Tests

Ch 93 LO Measurement 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM leads as shown in Figure 2-7• Turn mainframe power ON 2. Me

Strany 21 - Verification

Ch 93 GU Measurement 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM leads as shown in Figure 2-8• Turn mainframe power ON 2. Me

Strany 22 - Resistance Test

Example: ClosedChannel ResistanceTestThis example performs a Closed Channel Resistance Test to measureChannels 00 - 47, Channels 90-92 and Channel 93

Strany 23 - Verification Tests 2-5

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Strany 24 - 2-6 Verification Tests

300 PRINT " 5. Turn Mainframe power ON "310 PRINT " 5. Press Continue when ready to begin testing "320 PAUSE330 CLEA

Strany 25 - Verification Tests 2-7

670 PRINT TABXY(1,1),"Channels 90-92 ";Path$(K);"Measurements"680 PRINT TABXY(1,3),"Connect DMM Sense and Input

Strany 26 - 2-8 Verification Tests

1040 PAUSE1050 CLEAR SCREEN1060 OUTPUT @Mux;"CLOS (@"&VAL$(J)&")"1070 OUTPUT @Dmm;"TRIG SGL"1

Strany 27 - Verification Tests 2-9

1440 NEXT K1450 PRINT1460 PRINT "Channel 93 Contact Resistance"1470 PRINT1480 FOR K=0 TO 21490 PRINT "Channel 93 ";Path

Strany 28 - 2-10 Verification Tests

Test 2-2: DCIsolation TestThis test verifies that sufficient DC isolation exists at various points on themultiplexer. DC Isolation is checked from HI

Strany 29

HI to GU Isolation 1. Make hardware connections as shown in Figure 2-112. Trigger the DMM with TRIG SGL 3. Record the DMM reading on Table 2-2 (HI

Strany 30 - 2-12 Verification Tests

Example:DC Isolation TestThis example performs DC Isolation Tests for HI to Chassis, HI to LO, andHI to GU (GUARD).10! RE-SAVE "DC_ISOL" 2

Strany 31 - Verification Tests 2-13

360 PAUSE 370 CLEAR SCREEN 380 PRINT TABXY(1,1),"DC Isolation Tests "390 PRINT TABXY(1,3),"HI to CHASSIS (Ohms)";Value(0) 400

Strany 32 - 2-14 Verification Tests

PerformanceTest RecordTable 2-2, Performance Test Record, is a form you can copy and use torecord performance verification test results for the multip

Strany 33 - Verification Tests 2-15

Table 2-2. Performance Test Record (Page 1 of 3)General InformationTest Facility:Name ____________________________________Address _____________

Strany 34 - Isolation Test

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Strany 35 - Verification Tests 2-17

Table 2-2. Performance Test Record (Page 2 of 3)Test No/DescriptionMinimum*Value Measured Value (V)MaximumValue **MeasUncert Test AccRatio (TAR)2-1

Strany 36 - 2-18 Verification Tests

Table 2-2. Performance Test Record (Page 3 of 3)Test No/Description Minimum*ValueMeasured Value (V) MaximumValue **MeasUncert Test AccRatio (TAR)2

Strany 38 - Test Record

3Replaceable PartsIntroduction This chapter contains information to order replaceable parts for AgilentE1346A Single Ended Relay Multiplexers with ser

Strany 39 - Verification Tests 2-21

Table 3-2. Agilent E1346A Replaceable PartsReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberASSEMBLIES/CABLES/MANUALS(

Strany 40 - 2-22 Verification Tests

ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberA1CR2 8150-4086 1 WIRE 22AWG-WHITE TEFLON 1X22 105C 28480 8150-4086A1

Strany 41 - Verification Tests 2-23

ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberA1U17-U18 1820-3631 2 IC COMPARATOR CMOS/HCT MAGNITUDE 8-BIT 27014 MM

Strany 42 - 2-24 Verification Tests

ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberTERMINAL MODULE(See Figure 3-2)A3 E1346-66510 1 THERMNL MODULE PC 284

Strany 43 - Replaceable Parts

Table 3-4. Agilent E1346A Code List of Manufacturers Mfr.CodeManufacturer’sNameManufacturer’sAddressZipCode01295 TEXAS INSTRUMENTS INC DALLAS TX US

Strany 44 - 3-2 Replaceable Parts

Mechanical Parts LocatorsFigures 3-1 and 3-2 show locations of selected mechanical parts for theAgilent E1346A Single Ended Relay Multiplexer. See the

Strany 45 - Replaceable Parts 3-3

Agilent 75000 Series B Service DocumentationSuggested Sequence to Use ManualsManual DescriptionsInstallation and Getting Started Guide.

Strany 46 - 3-4 Replaceable Parts

Figure 3-2. Terminal Block Mechanical Parts3-8 Replaceable Parts

Strany 47 - Replaceable Parts 3-5

4ServiceIntroduction This chapter contains service information for the Agilent E1346Amultiplexer. Also included are troubleshooting, repair, and maint

Strany 48 - 3-6 Replaceable Parts

Agilent E1346AMultiplexerDescriptionThe Agilent E1346A 48-Channel Single Ended Relay Multiplexer consistsof an E1346A component assembly (E1346-66201)

Strany 49 - Locators

Figure 4-1. Agilent E1346A Block DiagramService 4-3

Strany 50 - 3-8 Replaceable Parts

Troubleshooting TechniquesTo troubleshoot a relay multiplexer problem you must first identify theproblem and then isolate the cause of the problem to

Strany 51 - Required

NOTESee the Agilent E1346A Single Ended Relay Multiplexer User’s Manual forinformation on logical address and IRQ settings. If there are no apparent

Strany 52 - Agilent E1346A

Table 4-3. TroubleshootingChannel Number Channel Relay Tree Relay Channel RelayDriver/Decoder00 K0K90U1, U3801 K102 K203 K304 K405 K506 K607 K708 K0K9

Strany 53 - Service 4-3

Repair and Maintenance GuidelinesThis section provides guidelines for repairing and maintaining the AgilentE1346A Single Ended Relay Multiplexer inclu

Strany 54 - Techniques

Component Locators andSchematicDiagrams Table 4-4 lists Component Locator Diagrams and Schematic Diagrams forthe Agilent E1346A relay multiplexer. Tab

Strany 55 - Testing the

Appendix AVerification Tests - C ProgramsFunctionalVerificationTestThis program is designed to do the Functional Verification Test foundin Chapter 2 -

Strany 56 - 4-6 Service

What’s in this ManualManual Overview This manual shows how to service the Agilent E1346A Single Ended Relay Multiplexer. Consult theAgilent E1346A Us

Strany 57 - Maintenance

PerformanceVerificationTestThis program is designed to do the Performance Verification Test foundin Chapter 2 - Verification Tests.NOTEThese programs

Strany 58 - Diagrams

/*---------------Measure Protection Resistors ---------------* / printf ("\ n\ nMeasure Protection Resistors R38 - R43"); printf ("\

Strany 59 - Functional

if (result[i] < 0) result[i] = 0;if (result[i] > 2.0) printf ("\n*** Resistance for Channel %u HI path is 2.0 Ohms*** %If", i, resu

Strany 60

for (i = 0; i <= 2; i++){iprintf (id, "CLOS (@°/0u)\ n", j); iprintf (dm, "TRIG SGL\ n");iscanf (dm, "%lf", &valu

Strany 61

iscanf (dm, "%t", cr);resultl[1][3] = valuel[1][3] - R[39] - R[42]; iprintf (id, "open (@%u)\ n", j);if (result1[1][3] < 0) r

Strany 62

Example: DCIsolation TestThis example performs DC Isolation Tests for HI to Chassis, HI to LO,and HI to GU (Guard)./* DC Isolation Test E1346A */# inc

Strany 63

/*...HI to GUARD...*/printf ("\ n\ nConnect DMM HI and LO to E1346A COMMON HI and COMMON G"); getchar ();iprintf (dm, "TR

Strany 64

ContentsChapter 1 - General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 65 - Example: DC

Chapter 3 - Replaceable PartsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 66

1General InformationIntroduction This manual contains information required to test, troubleshoot, and repairthe Agilent E1346A Single Ended Relay Mult