Agilent Technologies E1401-90493 Uživatelský manuál Strana 264

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 642
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 263
264 Chapter 2
Performance Verification Tests
31. Spurious Responses: Agilent E4402B and E4403B
Adapter, Type-N (m) to BNC (f)
Adapter, Type-N (m) to Type-N (m)
Adapter, SMA (f) to BNC (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Procedure
This performance test consists of two parts:
Part 1: Third Order Intermodulation Distortion
Part 2: Second Harmonic Distortion
Perform Part 1 before Part 2.
Part 1: Third Order Intermodulation Distortion
1. Zero and calibrate the power meter and RF power sensor in log mode
(power reads out in dBm), as described in the power meter operation
manual.
2. Connect the equipment as shown in Figure 2-47 with the output of
the directional bridge connected to the power sensor.
Figure 2-47 Third Order Intermodulation Distortion Test Setup
Zobrazit stránku 263
1 2 ... 259 260 261 262 263 264 265 266 267 268 269 ... 641 642

Komentáře k této Příručce

Žádné komentáře