Chapter 3 51
Operation
11974Q Rear-Panel Features
Operation
11974Q Rear-Panel Features
Figure 3-2. 11974Q Rear Panel
POWER SUPPLY
receives power from the 11974-60028 Preselector Power Supply.
a
a. Refer to Table 1-2 on page -19 for the ratings of these input and output connec-
tors.
INSTRUMENT
SELECT
switches allow the 11974 to properly scale the tune ramp of the specified
spectrum analyzer.
IF OUT requires an SMA type cable and supplies the IF INPUT for the spectrum
analyzer.
a
TUNE IN requires a BNC type cable and receives the TUNE + SPAN sweep output
from the spectrum analyzer.
a
LO IN requires an SMA type cable and receives the first LO output from the
spectrum analyzer.
a
33 GHz SET is the low-end of the band adjustment for mating the 11974 to the
spectrum analyzer. This adjustment frequency directly corresponds to the
low-end frequency of each band.
50 GHz SET is the high-end of the band adjustment for mating the 11974 to the
spectrum analyzer. This adjustment frequency directly corresponds to the
high-end frequency of each band.
LED SET
are the LEDs used to adjust the low-end and high-end of each band so that
the 11974 is mated properly to the spectrum analyzer.