Agilent Technologies HP8648A Uživatelský manuál Strana 293

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 320
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 292
RF Analyzer Specifications
292
RF Analyzer Specifications
RF Power Measurement
1
1. To achieve the specified accuracy when measuring power at the RF IN/OUT port, the internal signal generator
level must be 60 dB below the measured power or less than -20 dBm at the DUPLEX port.
Standard
Frequency Range: 400 kHz to 1 GHz
Measurement Range:
1 mW to 60 W continuous
100 W for 10 seconds/minute (measured at RF IN/OUT connector)
Accuracy: ±5% of reading ±0.01 mW (for temp 25° C ±10° C)
Accuracy: ±10% of reading for operating temperature range.
SWR: <1.5:1
Resolution:
Power <10 W: 1 mW
Power 10 W: 10 mW
Option 007
Frequency Range: 400 kHz to 1 GHz
Measurement Range:
40 µW to 2.4 W continuous
4.0 W for 10 seconds/minute
Accuracy: ±5% of reading ±400 nW (for temp 25° C ±10° C)
SWR: < 1.5:1
Resolution:
P < 400 mW: 40 µW
P 400 mW: 400 µW
Zobrazit stránku 292
1 2 ... 288 289 290 291 292 293 294 295 296 297 298 ... 319 320

Komentáře k této Příručce

Žádné komentáře