Agilent Technologies InfiniiVision 5000 Series Uživatelská příručka Strana 112

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 296
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 111
112 5000 Series Oscilloscope User’s Guide
3 Triggering the Oscilloscope
To use I
2
C triggering
An I
2
C (Inter-IC bus) trigger setup consists of connecting the
oscilloscope to the serial data (SDA) line and the serial clock
(SCL) line, then triggering on a stop/start condition, a restart, a
missing acknowledge, an EEPROM data read, or on a read/write
frame with a specific device address and data value.
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until I
2
C is displayed in the Trigger
softkey, then press the Settings softkey to display the I
2
C
trigger menu.
NOTE
For I
2
C decode setup see page 217.
Signals
menu
Trigger on:
condition
Address
Data 2
value
Trigger level
or threshold
Currently selected Clock
or Data channel
I
2
C trigger
Return to
previous menu
Data
value
Zobrazit stránku 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 295 296

Komentáře k této Příručce

Žádné komentáře