8 Acquisition Modes
274 InfiniiVision Oscilloscopes User’s Guide
Saving data from segmented memory
You can save either the currently displayed segment (Save Segment - Current), or all
segments (Save Segment - All) in the following data formats: CSV, ASCII XY, and BIN. Be
sure to set the Length control to capture enough points to accurately represent the
captured data. When the oscilloscope is busy saving multiple segments, progress is
displayed in the upper right area of the display.
To select the current segment or all segments, press
[Save/Recall]&Save&Settings&Save Seg. A USB storage device must be connected
to the oscilloscope to enable this Save option. You can also save the screen image in
BMP or PNG formats. For more information see “Save Segment” on page 253.
Select
segment to
display
Select
number of
segments
Choose
positioning
mode
Displayed segment
number and number of
segments acquired
Time (in seconds)
since the first trigger
Sample rate
Switch
segmented
on/off