N5230A/CCorrected System Performance with Type-N Connectors
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
85032B Calibration Kit
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
Table 13. 85032B Calibration Kit
N5230A/C- Option 020 (Standard Test Set and Standard Power Range)
Applies to the, N5230A/C Option 020 analyzers, 85032B (Type-N) calibration kit, 85132F flexible test port cable set with 85130C adapter
set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
0.300 to
1 MHz
1 to
10 MHz
10 to
45 MHz
45 MHz to
6 GHz
Directivity 50 50 47 40
Source Match 42 42 37 31
Load Match 50 41 47 38
Reflection Tracking
±0.009
+0.01/°C
±0.009
+0.01/°C
±0.019
+0.01/°C
±0.069
+0.02/°C
Transmission Tracking
±0.013
+0.01/°C
±0.007
+0.01/°C
±0.021
+0.01/°C
±0.101
+0.02/°C