User’s Guide 73
U3042AE12 Operational Check
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 24 on
page 66. This can be done using Limit Lines (press Trace > Limit Test) or Marker. The
trace should be above the limit. PASS will be displayed on the screen if the limit lines
are used.
11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.16” (16-Port).
Figure 56 Option 001 or 002 Reflection Tracking Trace (Port 1-16) with
N5230A/C
Figure 57 Standard Reflection Tracking Trace (Port 1-16) with N5230A/C
Komentáře k této Příručce