436 Chapter 16
GPIB Command Reference
SEGMNUM
SEGMNUM
Syntax SEGMNUM <numeric>
SEGMNUM?
Description For the active trace (set with the “TRAC” command on page 457), if the search range is set
to a segment for the list sweep (specified to SEGMENT with the “SEARNG” command on
page 432), sets a segment as the search range.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Search] - SEARCH RANGE MENU - SEGMENT NUMBER
SETCDATE
Syntax SETCDATE <year>,<month>,<day>
SETCDATE?
Description Sets the date of the built-in clock of the 4294A.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {year},{month},{day}<newline><^END>
Corresponding
key
[System] - CLOCK - DATE
<numeric>
Description Segment number of the search range
Range 1 to the number of segments in the list sweep table
Initial value 1
Resolution 1
<year> <month> <day>
Description Year Month Day
Range 1900 to 2099 1 to 12 1 to 31
Resolution 1 1 1