Agilent TechnologiesAgilent U3606A Multimeter|DC Power SupplyUser’s and Service Guide

Strany 2

X U3606A User’s and Service GuideDeclaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available on the Web site.

Strany 3 - Safety Symbols

3 DC Power Supply Operation Protection Functions80 U3606A User’s and Service GuideResetting the OVP conditionUse one of the following methods to rese

Strany 4 - General Safety Information

DC Power Supply Operation 3 Protection FunctionsU3606A User’s and Service Guide 81Overcurrent protection (OCP)In constant voltage mode, the U3606A reg

Strany 5 - Environmental Conditions

3 DC Power Supply Operation Protection Functions82 U3606A User’s and Service GuideChecking the OCP operationTo check OCP operation, raise the regulat

Strany 6 - Regulatory Markings

DC Power Supply Operation 3 Protection FunctionsU3606A User’s and Service Guide 83Resetting the OCP conditionUse one of the following methods to reset

Strany 7

3 DC Power Supply Operation Protection Functions84 U3606A User’s and Service GuideOvervoltage limit (OV)The overvoltage limit prevents the output vol

Strany 8 - In This Guide…

DC Power Supply Operation 3 Protection FunctionsU3606A User’s and Service Guide 85NOTE• The OV limit feature is always enabled by default whenever you

Strany 9

3 DC Power Supply Operation Protection Functions86 U3606A User’s and Service GuideOvercurrent limit (OC)The overcurrent limit prevents the output cur

Strany 10

DC Power Supply Operation 3 Protection FunctionsU3606A User’s and Service Guide 87NOTE• The OC limit feature is always enabled by default whenever you

Strany 11 - Table of Contents

3 DC Power Supply Operation Square-wave Output88 U3606A User’s and Service GuideSquare-wave OutputSquare- wave output is a unique function for many a

Strany 12 - 3 DC Power Supply Operation

DC Power Supply Operation 3 Square-wave OutputU3606A User’s and Service Guide 89Setting the amplitude1 To set the square- wave amplitude, press unti

Strany 13 - 4 System Related Operation

Table of ContentsU3606A User’s and Service Guide XITable of Contents1 Getting StartedIntroduction 2Measurement features 2Output features

Strany 14 - 6 Calibration Procedures

3 DC Power Supply Operation Square-wave Output90 U3606A User’s and Service GuideSetting the frequency1 To set the square- wave frequency, press unt

Strany 15 - 7 Disassembly and Repair

DC Power Supply Operation 3 Square-wave OutputU3606A User’s and Service Guide 91Setting the duty cycle1 To set the square- wave duty cycle, press un

Strany 16 - 9 List of Error Messages

3 DC Power Supply Operation Square-wave Output92 U3606A User’s and Service GuideSetting the pulse width1 To set the square- wave pulse width, press

Strany 17 - List of Figures

DC Power Supply Operation 3 Sweep FunctionsU3606A User’s and Service Guide 93Sweep FunctionsThe U3606A Multimeter|DC Power Supply is equipped with ram

Strany 18

3 DC Power Supply Operation Sweep Functions94 U3606A User’s and Service GuideGenerating a ramp signal1 Select your desired output mode by pressing Vo

Strany 19 - List of Tables

DC Power Supply Operation 3 Sweep FunctionsU3606A User’s and Service Guide 95Scan signalA typical scan signal length is based on the following paramet

Strany 20

3 DC Power Supply Operation Sweep Functions96 U3606A User’s and Service GuideGenerating a scan signal1 Select your desired output mode by pressing Vo

Strany 21 - Getting Started

DC Power Supply Operation 3 Selecting a RangeU3606A User’s and Service Guide 97Selecting a RangeThe U3606A Multimeter|DC Power Supply has the function

Strany 22 - Introduction

3 DC Power Supply Operation Enabling the Output98 U3606A User’s and Service GuideEnabling the OutputThe key controls turn the U3606A output on or o

Strany 23 - Output features

DC Power Supply Operation 3 Remote SensingU3606A User’s and Service Guide 99Remote SensingRemote sensing is used to maintain regulation at the load an

Strany 24 - Initial Inspection

Table of Contents XII U3606A User’s and Service GuidePerforming diode tests 51Selecting a Range 54Setting the Resolution 55Math Operati

Strany 25

3 DC Power Supply Operation Remote Sensing100 U3606A User’s and Service GuideRemote sensing connectionsRemote sensing requires connecting the load le

Strany 26 - Adjusting the Handle

DC Power Supply Operation 3 Remote SensingU3606A User’s and Service Guide 101Connecting the load leads to the rear terminal block1 Turn off the power.

Strany 27 - To Rack Mount the Instrument

3 DC Power Supply Operation Remote Sensing102 U3606A User’s and Service Guide4 Loosen the top screws with a Phillips screw driver and connect the rea

Strany 28 - Stacking the Instrument

DC Power Supply Operation 3 Remote SensingU3606A User’s and Service Guide 103Do not loosen the top screws of the rear output terminal block at a tilte

Strany 29 - Product at a Glance

3 DC Power Supply Operation Remote Sensing104 U3606A User’s and Service GuideEnable remote sensingTo configure the U3606A for remote sensing:1 Turn o

Strany 30 - The front panel at a glance

DC Power Supply Operation 3 Remote SensingU3606A User’s and Service Guide 105StabilityUsing remote sensing under certain combinations of load lead len

Strany 31 - The display at a glance

3 DC Power Supply Operation Remote Sensing106 U3606A User’s and Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 32 - 1Getting Started

107U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies4 System Related OperationUsing the Utility Menu 108Changing confi

Strany 33 - Getting Started 1

4 System Related Operation Using the Utility Menu108 U3606A User’s and Service GuideUsing the Utility MenuThe utility menu allows you to customize a

Strany 34 - The keypad at a glance

System Related Operation 4 Using the Utility MenuU3606A User’s and Service Guide 109Changing configurable settingsAll items in the utility menu are co

Strany 35

Table of ContentsU3606A User’s and Service Guide XIIIRemote Sensing 994 System Related OperationUsing the Utility Menu 108Changing configura

Strany 36

4 System Related Operation Using the Utility Menu110 U3606A User’s and Service GuideEditing valuesSome items in the utility menu allows you to select

Strany 37

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 111Utility Menu SummaryThe utility menu items are summarized in the tab

Strany 38

4 System Related Operation Utility Menu Summary112 U3606A User’s and Service GuideIObGPib U-CdC U-tMC• Select “GPib”, “U-tMC”, or “U-CdC” as the desi

Strany 39 - The terminals at a glance

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 113SCAnCV CC• Press Voltage to select scan setup for CV.• Press Current

Strany 40

4 System Related Operation Utility Menu Summary114 U3606A User’s and Service GuideReading error messagesThe following procedure shows you how to read

Strany 41

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1157 Repeat step 6 for all errors in the error queue.8 After reading a

Strany 42 - The rear panel at a glance

4 System Related Operation Utility Menu Summary116 U3606A User’s and Service GuideAdjusting the display brightnessThe following procedure shows you h

Strany 43 - Digital Multimeter Operation

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1174 Press or to change the power-on state. a Select “rESEt” to aut

Strany 44 - Making Measurements

4 System Related Operation Utility Menu Summary118 U3606A User’s and Service Guide• When a continuity measurement is less than or equal to the contin

Strany 45 - Making the connections

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1194 Press or to change the remote interface connection (GPib, U- C

Strany 46 - Measuring DC voltage

Table of Contents XIV U3606A User’s and Service GuideUsing an electronic load 143Connecting the current monitoring resistor 143Test Consider

Strany 47 - Measuring AC voltage

4 System Related Operation Utility Menu Summary120 U3606A User’s and Service GuidePerforming a self-testA power- on self- test occurs immediately whe

Strany 48 - Measuring AC+DC voltage

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 121If the power- on or complete self- test fails, the Error annunciator

Strany 49

4 System Related Operation Utility Menu Summary122 U3606A User’s and Service GuideSetting the output protection stateThe OVP and OCP protection featu

Strany 50 - Measuring DC current

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1235 Press Shift > Save to save or press Shift > Exit to exit th

Strany 51 - Measuring AC (rms) current

4 System Related Operation Utility Menu Summary124 U3606A User’s and Service Guide4 Press Voltage or Current if you wish to configure the ramp signa

Strany 52 - Measuring AC+DC current

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1258 After configuring the ramp signal parameters, press Shift > Sav

Strany 53

4 System Related Operation Utility Menu Summary126 U3606A User’s and Service Guide5 Select an appropriate output range by pressing Shift > Range.

Strany 54 - Measuring resistance

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 127Setting the smooth functionThe smooth function is a special short- t

Strany 55

4 System Related Operation Utility Menu Summary128 U3606A User’s and Service Guide4 Press or to enable or disable the smooth function, or “on” or

Strany 56

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 1296 To set the number of points, press until the cursor position is

Strany 57 - Measuring low-resistance

Table of ContentsU3606A User’s and Service Guide XVResetting the security code to the factory default 183Calibration Count 186Calibration Me

Strany 58

4 System Related Operation Utility Menu Summary130 U3606A User’s and Service GuideEnable refresh holdThe following procedure shows you how to enable

Strany 59

System Related Operation 4 Utility Menu SummaryU3606A User’s and Service Guide 131Enable data holdThe following procedure shows you how to enable the

Strany 60

4 System Related Operation Storing and Recalling Instrument States132 U3606A User’s and Service GuideStoring and Recalling Instrument StatesYou can s

Strany 61

System Related Operation 4 Storing and Recalling Instrument StatesU3606A User’s and Service Guide 133Recalling a stored stateTo recall an instrument s

Strany 62

4 System Related Operation Remote Operation134 U3606A User’s and Service GuideRemote OperationThe U3606A is shipped with both a GPIB (IEEE- 488) inte

Strany 63

System Related Operation 4 Remote OperationU3606A User’s and Service Guide 135Configuring and connecting the GPIB interfaceThe GPIB (IEEE- 488) connec

Strany 64

4 System Related Operation Remote Operation136 U3606A User’s and Service GuideConfiguring and connecting the USB interfaceSelect the appropriate USB

Strany 65

System Related Operation 4 Remote OperationU3606A User’s and Service Guide 137SCPI commandsThe U3606A complies with the syntax rules and conventions o

Strany 66

4 System Related Operation Remote Operation138 U3606A User’s and Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 67

139U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies5 Verification and Performance TestsRecommended Test Equipment 140G

Strany 68 - Measuring capacitance

Table of Contents XVI U3606A User’s and Service GuideDigital Multimeter Specifications 242Specification assumptions 242DC specifications

Strany 69 - Performing continuity tests

5 Verification and Performance Tests Recommended Test Equipment140 U3606A User’s and Service GuideRecommended Test EquipmentThe test equipment recomm

Strany 70 - Testing continuity

Verification and Performance Tests 5 Recommended Test EquipmentU3606A User’s and Service Guide 141Constant voltage source effect (line regulation)Digi

Strany 71 - Performing diode tests

5 Verification and Performance Tests Recommended Test Equipment142 U3606A User’s and Service GuideConstant current source effect (line regulation)Dig

Strany 72 - Checking diodes

Verification and Performance Tests 5 Recommended Test EquipmentU3606A User’s and Service Guide 143General measurement techniquesA measurement made acr

Strany 73

5 Verification and Performance Tests Test Considerations144 U3606A User’s and Service GuideTest ConsiderationsErrors may be induced by AC signals pre

Strany 74

Verification and Performance Tests 5 Input ConnectionsU3606A User’s and Service Guide 145Input ConnectionsTest connections to the instrument are best

Strany 75 - Setting the Resolution

5 Verification and Performance Tests Input Connections146 U3606A User’s and Service GuideGain verification test setupFigure 5-2 Test setup for DC vol

Strany 76 - Math Operations

Verification and Performance Tests 5 Input ConnectionsU3606A User’s and Service Guide 147Figure 5-4 Test setup for frequency gain verificationOutput v

Strany 77 - Enabling the null operation

5 Verification and Performance Tests Input Connections148 U3606A User’s and Service GuideFigure 5-6 Test setup for CV load and line regulation verifi

Strany 78 - Editing the null value

Verification and Performance Tests 5 Input ConnectionsU3606A User’s and Service Guide 149Figure 5-8 Test setup for load transient response time verifi

Strany 79 - Enabling the dBm operation

List of FiguresU3606A User’s and Service Guide XVIIList of FiguresFigure 1-1 U3606A handle adjustments 6Figure 1-2 U3606A stacking directions

Strany 80 - Enabling the dB operation

5 Verification and Performance Tests Input Connections150 U3606A User’s and Service GuideFigure 5-10 Test setup for CC line and load regulation verif

Strany 81 - Editing the relative value

Verification and Performance Tests 5 Input ConnectionsU3606A User’s and Service Guide 151Figure 5-12 Test setup for square-wave output verification432

Strany 82 - Enabling the MinMax operation

5 Verification and Performance Tests Verification and Performance Tests Overview152 U3606A User’s and Service GuideVerification and Performance Tests

Strany 83

Verification and Performance Tests 5 Self-testU3606A User’s and Service Guide 153Self-testA brief power- on self- test occurs automatically whenever y

Strany 84 - Enabling the limit operation

5 Verification and Performance Tests Performance Verification Tests154 U3606A User’s and Service GuidePerformance Verification TestsThe performance v

Strany 85

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 155Table 5 - 2 Zero offset verification testInput F

Strany 86

5 Verification and Performance Tests Performance Verification Tests156 U3606A User’s and Service GuideGain verification testThis test checks the full

Strany 87

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 157DC current gain verification test1 Connect the c

Strany 88 - Enabling the hold operation

5 Verification and Performance Tests Performance Verification Tests158 U3606A User’s and Service GuideAC current verification test1 Connect the calib

Strany 89 - Triggering the Multimeter

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 159Resistance gain verification test1 Connect the c

Strany 90 - Front panel triggering

List of Figures XVIII U3606A User’s and Service GuideFigure 5-10 Test setup for CC line and load regulation verification 150Figure 5-11 Test setu

Strany 91 - Remote interface triggering

5 Verification and Performance Tests Performance Verification Tests160 U3606A User’s and Service GuideFrequency gain verification test1 Connect the c

Strany 92

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 161Output verification testThis test verifies that

Strany 93 - DC Power Supply Operation

5 Verification and Performance Tests Performance Verification Tests162 U3606A User’s and Service GuideCV load effect (load regulation)This test measu

Strany 94 - Basic Operation

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 163CV source effect (line regulation)This test meas

Strany 95 - Supplying constant voltage

5 Verification and Performance Tests Performance Verification Tests164 U3606A User’s and Service GuideCV noise effectThis test measures the RMS or pe

Strany 96 - Constant current (CC) mode

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 1656 Configure the oscilloscope as follows:i Set th

Strany 97 - Supplying constant current

5 Verification and Performance Tests Performance Verification Tests166 U3606A User’s and Service Guide3 Set the output voltage to the full rated valu

Strany 98 - Protection Functions

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 167CC programming and readback accuracyThis test ve

Strany 99 - Checking the OVP operation

5 Verification and Performance Tests Performance Verification Tests168 U3606A User’s and Service Guide5 Operate the electronic load in constant resis

Strany 100 - Resetting the OVP condition

Verification and Performance Tests 5 Performance Verification TestsU3606A User’s and Service Guide 1696 Operate the electronic load in constant resist

Strany 101 - Overcurrent protection (OCP)

List of TablesU3606A User’s and Service Guide XIXList of TablesTab l e 1 - 1 U3606A display annunciators 11Tab l e 1 - 2 U3606A keypad function

Strany 102 - Checking the OCP operation

5 Verification and Performance Tests Performance Verification Tests170 U3606A User’s and Service Guide4 Turn on the instrument. Press Current to sele

Strany 103 - Resetting the OCP condition

Verification and Performance Tests 5 Additional Verification TestsU3606A User’s and Service Guide 171Additional Verification TestsOptional capacitance

Strany 104 - Overvoltage limit (OV)

5 Verification and Performance Tests Additional Verification Tests172 U3606A User’s and Service GuideOptional square-wave output verification testThi

Strany 105

Verification and Performance Tests 5 Additional Verification TestsU3606A User’s and Service Guide 173Square-wave frequency output verification test1 T

Strany 106 - Overcurrent limit (OC)

5 Verification and Performance Tests Additional Verification Tests174 U3606A User’s and Service GuideTable 5 - 18 Square-wave duty cycle output verif

Strany 107

175U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies6 Calibration ProceduresCalibration Overview 176Closed-case electr

Strany 108 - Square-wave Output

6 Calibration Procedures Calibration Overview176 U3606A User’s and Service GuideCalibration OverviewThis section contains procedures for verifying th

Strany 109 - Setting the amplitude

Calibration Procedures 6 Calibration OverviewU3606A User’s and Service Guide 177Calibration intervalThe instrument should be calibrated on a regular i

Strany 110 - Setting the frequency

6 Calibration Procedures Recommended Test Equipment178 U3606A User’s and Service GuideRecommended Test EquipmentThe test equipment recommended for th

Strany 111 - Setting the duty cycle

Calibration Procedures 6 Calibration ProcessU3606A User’s and Service Guide 179Calibration ProcessThe test equipment recommended for the calibration p

Strany 112 - Setting the pulse width

II U3606A User’s and Service GuideNotices© Agilent Technologies, Inc., 2009–2014No part of this manual may be reproduced in any form or by any means (

Strany 113 - Sweep Functions

List of Tables XX U3606A User’s and Service Guidetest 167Table 5-13 Constant current load effect verification test 168Table 5-14 Constant curr

Strany 114 - Generating a ramp signal

6 Calibration Procedures Calibration Security180 U3606A User’s and Service GuideCalibration SecurityThis feature allows you to enter a security code

Strany 115 - Scan signal

Calibration Procedures 6 Calibration SecurityU3606A User’s and Service Guide 181Unsecuring from the front panel Only four characters (the fourth thro

Strany 116 - Generating a scan signal

6 Calibration Procedures Calibration Security182 U3606A User’s and Service Guide4 Press or to increment or decrement the digit.5 Press when you

Strany 117 - Selecting a Range

Calibration Procedures 6 Calibration SecurityU3606A User’s and Service Guide 183Changing the calibration security codeThe security code can only be ch

Strany 118 - Enabling the Output

6 Calibration Procedures Calibration Security184 U3606A User’s and Service Guide1 Disconnect the power cord and all input connections.2 Disassemble t

Strany 119 - Remote Sensing

Calibration Procedures 6 Calibration SecurityU3606A User’s and Service Guide 1854 Apply power and turn on the instrument.5 The display will show the m

Strany 120 - Remote sensing connections

6 Calibration Procedures Calibration Count186 U3606A User’s and Service GuideCalibration CountThe calibration count feature provides an independent “

Strany 121

Calibration Procedures 6 Calibration MessageU3606A User’s and Service Guide 187Calibration MessageThe instrument allows you to store a message in cali

Strany 122

6 Calibration Procedures Using the Front Panel for Adjustments188 U3606A User’s and Service GuideUsing the Front Panel for AdjustmentsThis section de

Strany 123 - DC Power Supply Operation 3

Calibration Procedures 6 Using the Front Panel for AdjustmentsU3606A User’s and Service Guide 189When the cursor is positioned on a digit (the digit c

Strany 124 - Enable remote sensing

1U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies1 Getting StartedIntroduction 2Measurement features 2Output featu

Strany 125 - Output noise

6 Calibration Procedures Using the Front Panel for Adjustments190 U3606A User’s and Service GuideGeneral calibration procedureThe steps below shows h

Strany 126

Calibration Procedures 6 Using the Front Panel for AdjustmentsU3606A User’s and Service Guide 1913 Set up the indicated reference input and apply this

Strany 127 - System Related Operation

6 Calibration Procedures Using the Front Panel for Adjustments192 U3606A User’s and Service GuideIf the reading is out of the acceptable range, “FAiL

Strany 128 - Using the Utility Menu

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 193Adjustments proceduresYou will need a test input cable and connector

Strany 129

6 Calibration Procedures Adjustments procedures194 U3606A User’s and Service Guide3 Press Shift > Hold. The display “CALib” in the lower secondar

Strany 130 - Editing values

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 195Gain adjustmentsThe instrument calculates and stores gain correction

Strany 131 - Utility Menu Summary

6 Calibration Procedures Adjustments procedures196 U3606A User’s and Service GuideValid gain adjustment input values Gain adjustment can be accomplis

Strany 132

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 197DC voltage gain adjustment procedureReview the “Test Considerations”

Strany 133 - System Related Operation 4

6 Calibration Procedures Adjustments procedures198 U3606A User’s and Service Guide6 Enter the actual applied input (see “Entering adjustment values”

Strany 134 - Reading error messages

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 199AC voltage gain adjustment procedureReview the “Test Considerations”

Strany 135

1Getting Started Introduction2 U3606A User’s and Service GuideIntroductionThe Agilent U3606A Multimeter|DC Power Supply unit is a combination of a 5½

Strany 136 - Changing the power-on state

6 Calibration Procedures Adjustments procedures200 U3606A User’s and Service GuideFrequency gain adjustment procedureReview the “Test Considerations”

Strany 137 - Configuring the beeper

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 2017 Press to start the adjustment. The display “CALib” in the lower

Strany 138

6 Calibration Procedures Adjustments procedures202 U3606A User’s and Service Guide4 Use or to select the adjustment item.5 Apply the input signal

Strany 139

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 203DC current gain adjustment procedureReview the “Test Considerations”

Strany 140 - Performing a self-test

6 Calibration Procedures Adjustments procedures204 U3606A User’s and Service Guide• An adjustment failure is indicated by a long beep, the primary di

Strany 141

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 2056 Enter the actual applied input (see “Entering adjustment values” o

Strany 142

6 Calibration Procedures Adjustments procedures206 U3606A User’s and Service Guide4 Use or to select the adjustment item.5 Apply the input signal

Strany 143

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 20710.00 nF 10 nF100.0 nF 100 nF1.000 μF1 μF10.00 μF10 μF100.0 μF100 μF

Strany 144

6 Calibration Procedures Adjustments procedures208 U3606A User’s and Service GuideOutput adjustmentsThe instrument calculates and stores output corre

Strany 145

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 209CC output adjustment procedure — front output terminalsReview the “T

Strany 146

Getting Started 1 IntroductionU3606A User’s and Service Guide 3Output features• Single- output dual range — S1 (30 V/1 A) or S2 (8 V/3 A)• Constant vo

Strany 147 - Setting the smooth function

6 Calibration Procedures Adjustments procedures210 U3606A User’s and Service Guide• An adjustment failure is indicated by a long beep, the primary di

Strany 148

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 211CV output adjustment procedure — front output terminalsReview the “T

Strany 149

6 Calibration Procedures Adjustments procedures212 U3606A User’s and Service Guide6 Press to start the adjustment. The display “CALib” in the lower

Strany 150 - Enable refresh hold

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 2139 Verify the constant voltage adjustments using the “CV programming

Strany 151 - Enable data hold

6 Calibration Procedures Adjustments procedures214 U3606A User’s and Service Guide2 Press Shift > EXT to enable remote sensing. When the U3606A i

Strany 152 - Storing a state

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 2155 Verify the constant current adjustments using the “CC programming

Strany 153 - Recalling a stored state

6 Calibration Procedures Adjustments procedures216 U3606A User’s and Service Guide2 Press Shift > EXT to enable remote sensing. When the U3606A i

Strany 154 - Remote Operation

Calibration Procedures 6 Adjustments proceduresU3606A User’s and Service Guide 2175 Verify the constant voltage adjustments using the “CV programming

Strany 155 - GPIB address

6 Calibration Procedures Adjustments procedures218 U3606A User’s and Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 156

219U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies7 Disassembly and RepairOperating Checklist 220Cleaning 221Fuse

Strany 157 - SCPI commands

1Getting Started Initial Inspection4 U3606A User’s and Service GuideInitial InspectionWhen you receive your instrument, inspect the unit for any obvi

Strany 158

7 Disassembly and Repair Operating Checklist220 U3606A User’s and Service GuideOperating ChecklistBefore returning your instrument to Agilent for ser

Strany 159 - Agilent Technologies

Disassembly and Repair 7 CleaningU3606A User’s and Service Guide 221CleaningClean the outside of the instrument with a soft, lint- free, slightly damp

Strany 160 - Recommended Test Equipment

7 Disassembly and Repair Cleaning222 U3606A User’s and Service GuideS/B 5×20 mm fuse, Agilent part number A02- 62- 25026- 2U. If you are certain that

Strany 161

Disassembly and Repair 7 CleaningU3606A User’s and Service Guide 2232 Remove the fuse holder assembly from the rear panel.3 Remove the faulty power li

Strany 162

7 Disassembly and Repair Cleaning224 U3606A User’s and Service GuideTo replace the current input fuseThe current input terminal is fuse protected. Th

Strany 163 - Using an electronic load

Disassembly and Repair 7 CleaningU3606A User’s and Service Guide 2253 Remove the faulty current input fuse and replace a new current input fuse into t

Strany 164 - Test Considerations

7 Disassembly and Repair Electrostatic Discharge (ESD) Precautions226 U3606A User’s and Service GuideElectrostatic Discharge (ESD) PrecautionsElectri

Strany 165 - Input Connections

Disassembly and Repair 7 Mechanical DisassemblyU3606A User’s and Service Guide 227Mechanical DisassemblyFor procedures in this section, the following

Strany 166 - Alternate connection path

7 Disassembly and Repair Mechanical Disassembly228 U3606A User’s and Service Guide3 Remove the instrument bumpers. Pull from a corner and stretch the

Strany 167

Disassembly and Repair 7 Mechanical DisassemblyU3606A User’s and Service Guide 2295 Remove the cover. Remove the screws in the bottom of the cover.6 S

Strany 168

Getting Started 1 Connecting Power to the InstrumentU3606A User’s and Service Guide 5Connecting Power to the InstrumentConnect the power cord to the A

Strany 169

7 Disassembly and Repair Mechanical Disassembly230 U3606A User’s and Service GuideFront panel removal7 Remove the push rod and disconnect the display

Strany 170

Disassembly and Repair 7 Mechanical DisassemblyU3606A User’s and Service Guide 231c Disconnect the individual wires from the top layer circuit board a

Strany 171

7 Disassembly and Repair Mechanical Disassembly232 U3606A User’s and Service Guide9 Disconnect the individual wires from the bottom layer circuit boa

Strany 172

Disassembly and Repair 7 Mechanical DisassemblyU3606A User’s and Service Guide 23311 You can now pry the side of the front panel from the chassis and

Strany 173 - Self-test

7 Disassembly and Repair Mechanical Disassembly234 U3606A User’s and Service Guide13 Remove the keypad and display assembly. The rubber keypad and VF

Strany 174 - Zero offset verification test

Disassembly and Repair 7 Replaceable PartsU3606A User’s and Service Guide 235Replaceable PartsThis section contains information for ordering replaceme

Strany 175 - Input Function Range

7 Disassembly and Repair Replaceable Parts236 U3606A User’s and Service GuideTo Order Replaceable PartsYou can order replaceable parts from Agilent u

Strany 176 - Gain verification test

Disassembly and Repair 7 Types of Services AvailableU3606A User’s and Service Guide 237Types of Services AvailableIf your instrument fails during the

Strany 177 - AC voltage verification test

7 Disassembly and Repair Repackaging for Shipment238 U3606A User’s and Service GuideRepackaging for ShipmentIf the unit is to be shipped to Agilent f

Strany 178 - AC current verification test

239U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies8 Characteristics and SpecificationsProduct Characteristics 240Dig

Strany 179

1Getting Started Adjusting the Handle6 U3606A User’s and Service GuideAdjusting the HandleTo adjust the handle, grasp the handle by the sides and pul

Strany 180

8 Characteristics and Specifications Product Characteristics240 U3606A User’s and Service GuideProduct CharacteristicsPOWER SUPPLY• Universal 100 Vac

Strany 181 - Output verification test

Characteristics and Specifications 8 Product CharacteristicsU3606A User’s and Service Guide 241REMOTE INTERFACE• GPIB IEEE-488 compatible• Full Speed

Strany 182

8 Characteristics and Specifications Digital Multimeter Specifications242 U3606A User’s and Service GuideDigital Multimeter SpecificationsSpecificati

Strany 183

Characteristics and Specifications 8 Digital Multimeter SpecificationsU3606A User’s and Service Guide 243DC current[3]10.0000 mA < 0.2 V 0.05 + 0.0

Strany 184 - CV noise effect

8 Characteristics and Specifications Digital Multimeter Specifications244 U3606A User’s and Service GuideCapacitance[7]1.000 nF0.75 µA current sourc

Strany 185 - Load transient response time

Characteristics and Specifications 8 Digital Multimeter SpecificationsU3606A User’s and Service Guide 245AC specificationsTable 8 - 2 AC accuracy spec

Strany 186

8 Characteristics and Specifications Digital Multimeter Specifications246 U3606A User’s and Service GuideFrequency specificationsTable 8 - 3 Frequenc

Strany 187 - 3 A 3 A ±0.0075 A

Characteristics and Specifications 8 Digital Multimeter SpecificationsU3606A User’s and Service Guide 247Duty cycle and pulse width specifications100

Strany 188

8 Characteristics and Specifications Digital Multimeter Specifications248 U3606A User’s and Service GuideOperating specificationsTable 8 - 7 Reading

Strany 189 - CC noise effect

Characteristics and Specifications 8 Digital Multimeter SpecificationsU3606A User’s and Service Guide 249Supplementary characteristicsWARNINGExceeding

Strany 190

Getting Started 1 To Rack Mount the InstrumentU3606A User’s and Service Guide 7To Rack Mount the InstrumentYou can mount the U3606A in a standard 19-

Strany 191 - Additional Verification Tests

8 Characteristics and Specifications Digital Multimeter Specifications250 U3606A User’s and Service GuideCrest factor: For < 5:1 errors included.

Strany 192 - 10000 μF 10000 μF±250μF

Characteristics and Specifications 8 Digital Multimeter SpecificationsU3606A User’s and Service Guide 251ResistanceMeasurement method: Two-wire, open-

Strany 193

8 Characteristics and Specifications Digital Multimeter Specifications252 U3606A User’s and Service GuideResponse time: Approximately 1 s for 100 μF

Strany 194

Characteristics and Specifications 8 DC Power Supply SpecificationsU3606A User’s and Service Guide 253DC Power Supply SpecificationsSafety considerati

Strany 195 - Calibration Procedures

8 Characteristics and Specifications DC Power Supply Specifications254 U3606A User’s and Service GuidePerformance specificationsTable 8 - 8 DC power

Strany 196 - Calibration Overview

Characteristics and Specifications 8 DC Power Supply SpecificationsU3606A User’s and Service Guide 255Supplementary characteristicsFront panel resolut

Strany 197 - Time required for calibration

8 Characteristics and Specifications DC Power Supply Specifications256 U3606A User’s and Service GuideSquare-wave output specificationsRemote sensing

Strany 198

Characteristics and Specifications 8 DC Power Supply SpecificationsU3606A User’s and Service Guide 257[2] For frequency signals greater than 100 Hz,

Strany 199 - Calibration Process

8 Characteristics and Specifications DC Power Supply Specifications258 U3606A User’s and Service GuideSweep specificationsTable 8 - 10 Scan output sp

Strany 200 - Calibration Security

259U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies9 List of Error MessagesError Messages 260Command errors 261Exe

Strany 201

1Getting Started Stacking the Instrument8 U3606A User’s and Service GuideStacking the InstrumentIf you have purchased more than one unit of the U3606

Strany 202

9 List of Error Messages Error Messages260 U3606A User’s and Service GuideError MessagesError messages are created once an erroneous condition is det

Strany 203

List of Error Messages 9 Error MessagesU3606A User’s and Service Guide 261Command errorsThe following table shows the list of command errors. These er

Strany 204

9 List of Error Messages Error Messages262 U3606A User’s and Service GuideExecution errorsThe following table shows the list of execution errors. The

Strany 205

List of Error Messages 9 Error MessagesU3606A User’s and Service Guide 263Internal errorsThe following table shows the list of internal errors.Query e

Strany 206 - Calibration Count

9 List of Error Messages Error Messages264 U3606A User’s and Service GuideSelf-test errorsThe following errors indicate failures that may occur durin

Strany 207 - Calibration Message

List of Error Messages 9 Error MessagesU3606A User’s and Service Guide 265Calibration errorsThe following errors indicate failures that may occur duri

Strany 208 - Entering adjustment values

9 List of Error Messages Error Messages266 U3606A User’s and Service Guide726 Cal RES open out of range742 Cal checksum failed, DCV corrections743 Ca

Strany 209 - Key Description

© Agilent Technologies, Inc., 2009–2014Tenth Edition, July 3, 2014U3606-90013www.agilent.comAgilent TechnologiesContact usTo obtain service, warranty,

Strany 210 - General calibration procedure

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 9Product at a GlanceProduct dimensionsFigure 1-3 U3606A dimensions105 mm329 mm255

Strany 211

U3606A User’s and Service Guide IIISafety SymbolsThe following symbols on the instrument and in the documentation indicate precautions which must be t

Strany 212

1Getting Started Product at a Glance10 U3606A User’s and Service GuideThe front panel at a glanceFigure 1-4 U3606A front panelItem Description1Source

Strany 213 - Adjustments procedures

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 11The display at a glanceFigure 1-5 VFD full display with all segments illuminate

Strany 214

1Getting Started Product at a Glance12 U3606A User’s and Service GuideAC measurement selectedAC+DC measurement selectedºC Celsius temperature unit (f

Strany 215 - Gain adjustments

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 13Store Store instrument state selectedRecall Recall instrument state selectedRam

Strany 216

1Getting Started Product at a Glance14 U3606A User’s and Service GuideThe keypad at a glanceFigure 1-6 U3606A keypad with both multimeter and source

Strany 217

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 15Table 1 - 2 U3606A keypad functions Key DescriptionSystem related operationPush

Strany 218

1Getting Started Product at a Glance16 U3606A User’s and Service GuidePress Shift > Recall to recall a previously stored instrument state. See “Re

Strany 219

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 17Press MinMax to store statistical data for the current readings. See “MinMax” o

Strany 220

1Getting Started Product at a Glance18 U3606A User’s and Service Guide• Press to select the square-wave output. Use the directional keys to set the

Strany 221 - 1 kHz 1 V

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 19The terminals at a glanceFigure 1-7 U3606A connector terminals[3] The continuit

Strany 222

IV U3606A User’s and Service GuideGeneral Safety InformationWARNING• Do not exceed any of the measurement limits defined in the specifications to avoi

Strany 223 - 100.000 Ω 100 Ω

1Getting Started Product at a Glance20 U3606A User’s and Service GuideTable 1 - 3 Input terminal connections for measurement functionsFunction Input

Strany 224

Getting Started 1 Product at a GlanceU3606A User’s and Service Guide 21Table 1 - 4 Output terminal connections for source functionsFunction Output ter

Strany 225 - 10.0000 mA, 1 kHz 10 mA 1 kHz

1Getting Started Product at a Glance22 U3606A User’s and Service GuideThe rear panel at a glanceFigure 1-8 U3606A rear panelItem Description1 AC powe

Strany 226

23U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies2 Digital Multimeter OperationMaking Measurements 24Performing volt

Strany 227 - Calibration Procedures 6

2 Digital Multimeter Operation Making Measurements24 U3606A User’s and Service GuideMaking MeasurementsThe following pages introduce the many types o

Strany 228 - Output adjustments

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 25Performing voltage measurementsMaking the connectionsConnect the t

Strany 229

2 Digital Multimeter Operation Making Measurements26 U3606A User’s and Service GuideMeasuring DC voltage1 Press to make DC voltage measurements. (T

Strany 230 - OUt-H 3 A

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 27Measuring AC voltage1 Press again (until the AC annunciator is s

Strany 231

2 Digital Multimeter Operation Making Measurements28 U3606A User’s and Service GuideMeasuring AC+DC voltageThe U3606A Multimeter|DC Power Supply incl

Strany 232

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 29Performing current measurementsMaking the connectionsConnect the t

Strany 233

U3606A User’s and Service Guide VEnvironmental ConditionsThis instrument is designed for indoor use and in an area with low condensation. The table be

Strany 234 - +S + + – – –S

2 Digital Multimeter Operation Making Measurements30 U3606A User’s and Service GuideMeasuring DC current1 Press to make DC current measurement. (Th

Strany 235

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 31Measuring AC (rms) current1 Press again (until the AC annunciato

Strany 236

2 Digital Multimeter Operation Making Measurements32 U3606A User’s and Service GuideMeasuring AC+DC currentThe U3606A Multimeter|DC Power Supply incl

Strany 237

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 33Performing resistance (2-wire) measurementsMaking the connectionsC

Strany 238

2 Digital Multimeter Operation Making Measurements34 U3606A User’s and Service GuideMeasuring resistance1 Press to make 2- wire resistance measurem

Strany 239 - Disassembly and Repair

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 35Performing low-resistance (4-wire) measurementsFor measuring low-

Strany 240 - Operating Checklist

2 Digital Multimeter Operation Making Measurements36 U3606A User’s and Service GuideMaking the connectionsConnect the test leads as shown below:Figur

Strany 241 - Fuse Replacement

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 37Measuring low-resistance1 Press Shift > Lo Ω to make 4- wire l

Strany 242

2 Digital Multimeter Operation Making Measurements38 U3606A User’s and Service GuidePerforming frequency, pulse width, and duty cycle measurementsThe

Strany 243

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 39Figure 2-6 Terminal connections for frequency, pulse width, and du

Strany 244

VI U3606A User’s and Service GuideRegulatory MarkingsThe CE mark is a registered trademarkof the European Community. This CEmark shows that the produc

Strany 245

2 Digital Multimeter Operation Making Measurements40 U3606A User’s and Service GuideMeasuring frequency — voltage pathThe range and resolution of the

Strany 246

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 41Measuring pulse width — voltage pathThe range and resolution of th

Strany 247 - Mechanical Disassembly

2 Digital Multimeter Operation Making Measurements42 U3606A User’s and Service GuideMeasuring duty cycle — voltage pathThe range and resolution of th

Strany 248

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 43Measuring frequency — current pathThe range and resolution of the

Strany 249 - Disassembly and Repair 7

2 Digital Multimeter Operation Making Measurements44 U3606A User’s and Service GuideMeasuring pulse width — current pathThe range and resolution of t

Strany 250 - Front panel removal

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 45Measuring duty cycle — current pathThe range and resolution of the

Strany 251

2 Digital Multimeter Operation Making Measurements46 U3606A User’s and Service GuidePerforming capacitance measurementsThe U3606A Multimeter|DC Power

Strany 252

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 47Making the connectionsConnect the test leads as shown below:Figure

Strany 253 - Front panel disassembly

2 Digital Multimeter Operation Making Measurements48 U3606A User’s and Service GuideMeasuring capacitance1 Press to make capacitance measurements.2

Strany 254

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 49Performing continuity testsMaking the connectionsConnect the test

Strany 255 - Replaceable Parts

U3606A User’s and Service Guide VIIWaste Electrical and Electronic Equipment (WEEE) Directive 2002/96/ECThis instrument complies with the WEEE Directi

Strany 256 - To Order Replaceable Parts

2 Digital Multimeter Operation Making Measurements50 U3606A User’s and Service GuideTesting continuity1 Press again (until the continuity annunciat

Strany 257 - Types of Services Available

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 51Performing diode testsMaking the connectionsConnect the test leads

Strany 258 - Repackaging for Shipment

2 Digital Multimeter Operation Making Measurements52 U3606A User’s and Service GuideChecking diodes1 Press again (until the diode annunciator is sh

Strany 259

Digital Multimeter Operation 2 Making MeasurementsU3606A User’s and Service Guide 535 Reverse the probes and measure the voltage across the diode agai

Strany 260 - Product Characteristics

2 Digital Multimeter Operation Selecting a Range54 U3606A User’s and Service GuideSelecting a RangeYou can allow the U3606A to automatically select t

Strany 261

Digital Multimeter Operation 2 Setting the ResolutionU3606A User’s and Service Guide 55Setting the ResolutionYou can select either 4½ digit or 5½ digi

Strany 262 - DC specifications

2 Digital Multimeter Operation Math Operations56 U3606A User’s and Service GuideMath OperationsThe U3606A Multimeter|DC Power Supply provides six mat

Strany 263

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 57NullWhen making null measurements, also called relative, each reading

Strany 264

2 Digital Multimeter Operation Math Operations58 U3606A User’s and Service GuideEditing the null value1 Press Null again to view and edit the null v

Strany 265 - AC specifications

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 59dBm measurementsThe logarithmic dBm (decibels relative to 1 mW) scale

Strany 266 - Frequency specifications

VIII U3606A User’s and Service GuideIn This Guide…1 Getting StartedThis chapter prepares the U3606A Multimeter|DC Power Supply for use, and contains a

Strany 267

2 Digital Multimeter Operation Math Operations60 U3606A User’s and Service GuidedB measurementsEach dB measurement is the difference between the inpu

Strany 268 - Operating specifications

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 61Editing the relative value1 Press dB again to view and edit the relat

Strany 269 - Supplementary characteristics

2 Digital Multimeter Operation Math Operations62 U3606A User’s and Service GuideMinMaxThe MinMax operation stores the minimum and maximum values, the

Strany 270

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 63Each time a new minimum, maximum, or average value is stored, the inst

Strany 271

2 Digital Multimeter Operation Math Operations64 U3606A User’s and Service GuideLimitThe limit test function enables you to perform pass or fail test

Strany 272

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 65The primary display briefly shows “PASS” when readings are within the

Strany 273 - Specifications assumptions

2 Digital Multimeter Operation Math Operations66 U3606A User’s and Service GuideEditing the upper and lower limit values1 Press Shift > Limit aga

Strany 274 - Performance specifications

Digital Multimeter Operation 2 Math OperationsU3606A User’s and Service Guide 67HoldThe refresh hold feature allows you to capture and hold a reading,

Strany 275

2 Digital Multimeter Operation Math Operations68 U3606A User’s and Service GuideEnabling the hold operationPress Hold to enable the hold operation.T

Strany 276 - Range S1 0.2 V

Digital Multimeter Operation 2 Triggering the MultimeterU3606A User’s and Service Guide 69Triggering the MultimeterThe U3606A Multimeter|DC Power Supp

Strany 277

U3606A User’s and Service Guide IX8 Characteristics and SpecificationsThis chapter specifies the characteristics, environmental conditions, and specif

Strany 278 - Sweep specifications

2 Digital Multimeter Operation Triggering the Multimeter70 U3606A User’s and Service GuideFront panel triggeringSingle triggeringThe U3606A takes one

Strany 279 - List of Error Messages

Digital Multimeter Operation 2 Triggering the MultimeterU3606A User’s and Service Guide 71Remote interface triggeringImmediate triggeringIn the immedi

Strany 280 - Error Messages

2 Digital Multimeter Operation Triggering the Multimeter72 U3606A User’s and Service GuideRemote interface operation:• The TRIGger:SOURce BUS comman

Strany 281 - Command errors

73U3606A Multimeter|DC Power SupplyUser’s and Service GuideAgilent Technologies3 DC Power Supply OperationBasic Operation 74Constant voltage (CV)

Strany 282 - Execution errors

3 DC Power Supply Operation Basic Operation74 U3606A User’s and Service GuideBasic OperationThe U3606A has two basic operating modes: constant voltag

Strany 283 - Device specific errors

DC Power Supply Operation 3 Basic OperationU3606A User’s and Service Guide 75Supplying constant voltage1 Press Voltage to select the constant voltage

Strany 284 - Self-test errors

3 DC Power Supply Operation Basic Operation76 U3606A User’s and Service GuideConstant current (CC) modeMaking the connectionsConnect the load as show

Strany 285 - Calibration errors

DC Power Supply Operation 3 Basic OperationU3606A User’s and Service Guide 77Supplying constant current1 Press Current to select the constant current

Strany 286

3 DC Power Supply Operation Protection Functions78 U3606A User’s and Service GuideProtection FunctionsOvervoltage protection (OVP)In constant current

Strany 287

DC Power Supply Operation 3 Protection FunctionsU3606A User’s and Service Guide 794 Press Shift > Save or Voltage (Protect) to save the changes. “