Agilent TechnologiesAgilent U1241B and U1242B Handheld Digital MultimetersUser’s and Service Guide
X U1241B/U1242B User’s and Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1241B/U1242B User’s and Service Guide XContentsContents1Getting Started1Introduction 2Checking the Shipping Contents 3The Front Panel at a Glance 4Ad
XI U1241B/U1242B User’s and Service GuideContents3 Default Setting Configurations 27 Setting Configurations 284 Service and Maintenance 31 General
ContentsU1241B/U1242B User’s and Service Guide XII6 Specifications and Characteristics 61DC Specifications 62AC Specifications 63Resistance Specificat
XIII U1241B/U1242B User’s and Service GuideContents
1Agilent TechnologiesU1241B and U1242B Handheld Digital MultimetersUser’s and Service Guide1Getting Started Introduction 2 Checking the Shipping Conte
2 U1241B/U1242B User’s and Service Guide1Getting Started IntroductionThe handheld digital multimeters’ key features are:• DC, AC voltage and current
Getting Started 1U1241B/U1242B User’s and Service Guide 3Checking the Shipping ContentsVerify that you have received the following items for the sta
4 U1241B/U1242B User’s and Service Guide1Getting Started The Front Panel at a GlanceFigure 1-1 Front panel of a U1241B and U1242B handheld digital mu
Getting Started 1U1241B/U1242B User’s and Service Guide 5The Annunciator at a GlanceTo view the full display, press and hold while turning the rot
II U1241B/U1242B User’s and Service GuideNotices© Agilent Technologies, Inc. 2009–2014No part of this manual may be reproduced in any form or by any
6 U1241B/U1242B User’s and Service Guide1Getting Started Analog Bar GraphWhen frequency is indicated on the primary display during voltage or current
Getting Started 1U1241B/U1242B User’s and Service Guide 7Table 1-4 Keypad descriptions and functionsTable 1-5 Features that can be accessed through
8 U1241B/U1242B User’s and Service Guide1Getting Started The Input Terminal at a GlanceFigure 1-5 Input terminal of a U1242B handheld digital multime
9Agilent TechnologiesU1241B and U1242B Handheld Digital MultimeterUser’s and Service Guide2
10 U1241B/U1242B User’s and Service Guide2 Features and Functions Measuring VoltageMeasuring DC Voltage
Features and Functions 2U1241B/U1242B User’s and Service Guide 11Measuring Current (< 440 mA)Measuring % Scale of 4 – 20 mA NOTEIf the measured v
12 U1241B/U1242B User’s and Service Guide2 Features and Functions Measuring Frequency Measuring Resistance and Testing ContinuityCAUTIONDisconnec
Features and Functions 2U1241B/U1242B User’s and Service Guide 13Testing Dio desNOTEPress button to select measurement range from 1 kW to 100 MW.C
14 U1241B/U1242B User’s and Service Guide2 Features and Functions Measuring CapacitanceMeasuring TemperatureCAUTIONDisconnect circuit power and d
Features and Functions 2U1241B/U1242B User’s and Service Guide 15If you are working in a varied environment, where ambient temperature is not consta
U1241B/U1242B User’s and Service Guide IIISafety SymbolsThe following symbols on the instrument and in the documentation indicate precautions which mu
16 U1241B/U1242B User’s and Service Guide2 Features and Functions Measuring Harmonic Ratio (U1242B)Using Switch CounterHarmonics ratio function i
Features and Functions 2U1241B/U1242B User’s and Service Guide 171 Remove the power on the contacts or switch before measured. 2 Press at positi
18 U1241B/U1242B User’s and Service Guide2 Features and Functions MinMax RecordingThe MinMax operation stores the maximum, minimum, and average i
Features and Functions 2U1241B/U1242B User’s and Service Guide 19Data Hold (Trigger Hold)The data hold function allows users to freeze the displayed
20 U1241B/U1242B User’s and Service Guide2 Features and Functions Null (Relative)The Null function subtracts a stored value from the present meas
Features and Functions 2U1241B/U1242B User’s and Service Guide 21Manual LoggingTo enable the Hand (manual) logging function, select the Hand logging
22 U1241B/U1242B User’s and Service Guide2 Features and Functions 3 Press (Log) for more than one second to exit this mode.Figure 2-6 Manual lo
Features and Functions 2U1241B/U1242B User’s and Service Guide 23Figure 2-7 Interval logging displayReviewing Logged Data1 Press (View) for more t
24 U1241B/U1242B User’s and Service Guide2 Features and Functions and display temperature T1, T2 and T1- T2 sequentially. 1 Press and hold (Scan
Features and Functions 2U1241B/U1242B User’s and Service Guide 25Alerts and Warning During Measurement Overload AlertThe multimeter provides an over
IV U1241B/U1242B User’s and Service GuideRegulatory MarkingsWaste Electrical and Electronic Equipment (WEEE) Directive 2002/96/ECThis instrument comp
26 U1241B/U1242B User’s and Service Guide2 Features and Functions THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
27Agilent TechnologiesU1241B and U1242B Handheld Digital MultimeterUser’s and Service Guide3 De
28 U1241B/U1242B User’s and Service Guide3 Default Setting ConfigurationsSetting Configurations1 Turn off the multimeter. 2 From OFF position, press
Default Setting Configurations 3U1241B/U1242B User’s and Service Guide 29Table 3-9 Available setting options in Setup modeMenu item Available setting
30 U1241B/U1242B User’s and Service Guide3 Default Setting Configurations[1] To activate the multimeter after it has auto power off, press any button
31Agilent TechnologiesU1241B and U1242B Handheld Digital MultimeterUser’s and Service Guide4 Service
32 U1241B/U1242B User’s and Service Guide4 Service and Maintenance General MaintenanceBesides the above hazard, dirt or moisture in the terminals can
Service and Maintenance 4U1241B/U1242B User’s and Service Guide 333 Lift and remove the battery cover.4 Replace the specified batteries, ensure t
34 U1241B/U1242B User’s and Service Guide4 Service and Maintenance 3 Gently remove the defective Fuse 1 by prying one end of the fuse and removing it
Service and Maintenance 4U1241B/U1242B User’s and Service Guide 357 Replace a new fuse of the same size and rating into the center of the fuse ho
U1241B/U1242B User’s and Service Guide VGeneral Safety InformationThe following general safety precautions must be observed during all phases of opera
36 U1241B/U1242B User’s and Service Guide4 Service and Maintenance When servicing, use the specified replacement parts only. The Table 4- 11 shows th
Service and Maintenance 4U1241B/U1242B User’s and Service Guide 37Returning Instrument for ServiceBefore shipping your instrument for repair or r
38 U1241B/U1242B User’s and Service Guide4 Service and Maintenance THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
39Agilent TechnologiesU1241B and U1242B Handheld Digital MultimeterUser’s and Service Guide5
40 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationCalibration OverviewClosed-case Electronic CalibrationThe multimeter feat
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 41Adjustment is RecommendedSpecifications are only guaranteed within the per
42 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationRecommended Test EquipmentThe test equipment recommended for the performa
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 43Basic Operating TestBasic Operating Test is used to test the basic operabi
44 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationInput-A Terminal TestThis test determines if the input warning of the A c
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 45Test ConsiderationsError may be induced by AC signals that present on the
VI U1241B/U1242B User’s and Service GuideWARNING• Inspect the test probes for damaged insulation or exposed metal, and check for continuity. Do not u
46 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationInput ConnectionsTest connections to the multimeter are best accomplished
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 47Table 5-2 Verification TestsStep Test Function Range 5520A Output Error
48 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationStep Test Function Range 5520A Output Error from nominal 1 yearU1241B U
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 49[1] The accuracy of 1 kW and 10 kW are specified after Math Null, that is
50 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationCalibration SecurityThe calibration security code prevents accidental or
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 51Changing the Multimeter Calibration Security Code from the Front Panel1 Wh
52 U1241B/U1242B User’s and Service Guide5 Performance Tests and Calibrationerror code E03. Ensure that the last 4 digit serial number is entered cor
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 53Adjustment Considerations1 Allow the multimeter to warm up and stabilize f
54 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationValid Adjustment Input Values Adjustment can be accomplished using the fo
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 55Calibration Adjustments1 Turn the rotary switch to Test Function position,
U1241B/U1242B User’s and Service Guide VIIEnvironment ConditionsThis instrument is designed for indoor use, in areas with low condensation and to be u
56 U1241B/U1242B User’s and Service Guide5 Performance Tests and Calibration9 Turn the rotary switch to the next function according to the Test Funct
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 57Step Test Function Cal Range Input Cal ItemU1241B U1242B3 Turn the rotary
58 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationStep Test Function Cal Range Input Cal ItemU1241B U1242B6 Turn the rotar
Performance Tests and Calibration 5U1241B/U1242B User’s and Service Guide 59Exiting Adjustment Mode1 Remove all shorting plugs and connectors from the
60 U1241B/U1242B User’s and Service Guide5 Performance Tests and CalibrationCalibration Count The multimeter provides the calibration count informati
61Agilent TechnologiesU1241B and U1242B Handheld Digital MultimeterUser’s and Service Guide6
62 U1241B/U1242B User’s and Service Guide6 Specifications and CharacteristicsDC SpecificationsTable 6-6 DC specifications with accuracy of ± (% of re
Specifications and Characteristics 6U1241B/U1242B User’s and Service Guide 63AC SpecificationsTable 6-7 AC specifications with accuracy of ± (% of re
64 U1241B/U1242B User’s and Service Guide6 Specifications and CharacteristicsResistance SpecificationsTable 6-8 Resistance specifications with accura
Specifications and Characteristics 6U1241B/U1242B User’s and Service Guide 65Temperature SpecificationsTable 6-10 Temperature specifications with acc
VIII U1241B/U1242B User’s and Service GuideDeclaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available on the
66 U1241B/U1242B User’s and Service Guide6 Specifications and CharacteristicsCapacitance SpecificationsTable 6-11 Capacitance specifications with acc
Specifications and Characteristics 6U1241B/U1242B User’s and Service Guide 67Frequency SpecificationsTable 6-13 Frequency specifications with accurac
68 U1241B/U1242B User’s and Service Guide6 Specifications and CharacteristicsOperating SpecificationsTable 6-14 Measuring rate of U1241B and U1242BFu
Specifications and Characteristics 6U1241B/U1242B User’s and Service Guide 69General CharacteristicsTable 6-15 General characteristics of U1241B and
70 U1241B/U1242B User’s and Service Guide6 Specifications and CharacteristicsWeight• 450 g with batteries• 400 g without batteriesWarranty • Please r
www.agilent.comAgilent Technologies© Agilent Technologies, Inc. 2009–2014Printed in MalaysiaTenth Edition, June 26, 2014U1241-90063Contact usTo obtain