Agilent TechnologiesAgilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service Guide
X Agilent U1251B/U1252B User’s and Service Guide
80 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Data Hold/Refresh Hold Mode1 Set OFF to enable the Data Hold mo
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 81Setting Data Logging Mode1 Set to “Hand” to enable the manual data logg
82 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Thermocouple Types (U1252B only)The thermocouple sensor types t
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 83Setting Reference Impedance for dBm MeasurementThe reference impedance
84 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Minimum Frequency MeasurementThe minimum frequency setup influe
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 85Setting Temperature UnitFour combination displays are available: • Cels
86 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingFigure 4-6 Temperature unit setupPress Press Press Press Press
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 87Setting Auto Power Saving Mode• The timer for APF (Auto Power OFF) can
88 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingFigure 4-7 Auto power saving setupPress Press Press Press Press
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 89Setting Percentage (%) Scale ReadoutThis setting converts the DC curren
Agilent U1251B/U1252B User’s and Service Guide XIII Contents1 Getting StartedIntroducing the U1251B/U1252B Handheld Digital Multimeter 2Check the ship
90 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Beep FrequencyThe driving frequency can be set to 2400, 1200, 6
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 91Setting Backlight Timer• The timer can be set to 1—99 seconds. The back
92 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Baud RateSelect the baud rate for remote control. The available
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 93Setting Parity CheckSelect the parity check for remote control. It can
94 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Data BitSelect the data bit for remote control. It can be set t
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 95Setting Echo Mode• Echo ON enables the return of characters to the PC i
96 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting Print ModePrint ON enables the printing of measured data to the
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 97Returning to Default Factory Settings• Press for more than 1 second t
98 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSetting the Battery VoltageBattery type for the multimeter can be set t
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 99Setting the FilterThis setting is used to filter AC signals in DC measu
XIV Agilent U1251B/U1252B User’s and Service Guide Measuring Capacitance 43Measuring Temperature 45Alerts and Warning During Measurement 49Overload al
100 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingTable 4-2 Filter defaultsParameter Firmware version Default settingFiL
101Agilent TechnologiesAgilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service Guide5MaintenanceIntroduction 102General maintenance 10
102 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceIntroductionGeneral maintenanceBesides the above hazard, dirt, or moisture in the termi
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 1031 At the rear panel, turn the screw on the battery cover from the LOCK to the OPEN
104 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceFigure 5-1 9 V rectangular batteryStorage considerationsAfter the first charge, it is r
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 105Charging the batteryA new rechargeable battery comes in a discharged condition and
106 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceUse the specified 24 V DC adaptor to charge the battery. Remember never to turn the rot
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 107Figure 5-2 Battery capacity display as trickle5 After pressing SHIFT or when the s
108 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceFigure 5-3 Self-testTable 5-2 Error messages Error Error messageOL1 No battery inside2
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 1096 The smart charging mode will start if the meter passes the self- test. The charg
Agilent U1251B/U1252B User’s and Service Guide XV Setting Minimum Frequency Measurement 84Setting Temperature Unit 85Setting Auto Power Saving Mode 87
110 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceFigure 5-4 Charging mode7 The charge end message (C–End) appears on the secondary displ
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 111Figure 5-6 Battery charging procedurePress to start chargeError-FullError-EmptyBat
112 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceFuse checking procedureIt is recommended that you check the fuses of the multimeter bef
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 1133 To test Fuse 1, place the tip of the test probe on the right half of input termi
114 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceReplacing the fuseFollow the procedures below to replace the fuse.1 Turn the meter off
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 115Figure 5-8 Fuse replacementTable 5-4 Fuse specificationsFuse Agilent part number R
116 Agilent U1251B/U1252B User’s and Service Guide5 MaintenanceTroubleshootingIf the instrument fails to operate, check the battery and the test leads
Maintenance 5Agilent U1251B/U1252B User’s and Service Guide 117Replaceable PartsThis section contains information for ordering replacement parts
118 Agilent U1251B/U1252B User’s and Service Guide5 Maintenance
119Agilent TechnologiesAgilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service Guide6Performance Tests and CalibrationCalibration Over
XVI Agilent U1251B/U1252B User’s and Service Guide 6 Performance Tests and CalibrationCalibration Overview 120Closed-case electronic calibration 120Ag
120 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationCalibration OverviewThis manual contains procedures to verify the
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 121Calibration intervalA 1- year interval is adequate for most appli
122 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationRecommended Test EquipmentThe test equipment recommended for the
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 123Basic Operating TestThe Basic Operating Test is to test the basic
124 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationCurrent terminal testThis test determines if the input warning of
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 125Charge terminal alert test This test determines if the charge ter
126 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationTest ConsiderationsLong test leads can also act as antennas which
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 127Calibration SecurityThe calibration security code prevents accide
128 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationPerformance Verification TestsUse the Performance Verification Te
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 129Perform the verification test steps in the following Table 6- 2:T
Agilent U1251B/U1252B User’s and Service Guide XVII Specification Assumptions 158Electrical Specifications 158DC Specifications 158AC Specifications 1
130 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationStep Test Func tion Range 5520A Output Error from nominal 1 ye
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 131Step Test Function Range 5520A Output Error from nominal 1 year
132 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationStep Test Function Range 5520A Output Error from nominal 1 year
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 133Step Test Function Range 5520A Output Error from nominal 1 year
134 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationNotes for verification test:1 The additional error to be added as
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 13510 10 A continuous, and additional 0.5% to specified accuracy whe
136 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationUnsecuring the instrument for calibrationBefore you can adjust th
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 137To change the instrument calibration security code from the front
138 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationTo unsecure the instrument without the security codeTo unsecure t
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 139Calibration Process To complete a full instrument calibration use
XVIII Agilent U1251B/U1252B User’s and Service Guide
140 Agilent U1251B/U1252B User’s and Service Guide6 Performance Tests and CalibrationUsing the front panel for adjustmentsThis section describes the p
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 141Adjustments ConsiderationYou will need a test input cable, a conn
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 142Valid adjustment input values Adjustment can be accomplished usin
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 143Adjustment ProcedureReview the sections “Test Considerations” on
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 144For an adjustment that is successful, the secondary display will
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 145NOTEFor serial numbers below MY51530001, the 10 kHz input frequen
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 146Step Test Function Cal Range Input Cal ItemU1251B U1252B3Press
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 147Step Test Func tion Cal Range Input Cal ItemU1251B U1252B5 Press
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 148Step Test Function Cal Range Input Cal ItemU1251B U1252B7 Turn th
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 149Step Test Function Cal Range Input Cal ItemU1251B U1252B11 Press
Agilent U1251B/U1252B User’s and Service Guide IX List of FiguresFigure 1-1 Tilt-stand at 60° 4Figure 1-2 Tilt-stand at 30° 4Figure 1-3 Tilt-stand at
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 150Finishing the Adjustment1 Remove all the shorting plugs and conne
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 1511 Press Adjustment Mode. The primary display shows the calibrat
Performance Tests and Calibration 6Agilent U1251B/U1252B User’s and Service Guide 152
153Agilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service GuideAgilent Technologies7SpecificationsProduct Characteristics 154Measurem
154 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsProduct CharacteristicsPOWER SUPPLYBattery type: • 9 V size Ni-MH rechargeable batte
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 155SAFETY COMPLIANCE• EN/IEC 61010-1:2001• ANSI/UL 61010-1:2004• CAN/CSA-C22.2 No. 6101
156 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsWARRANTYPlease refer to http://www.agilent.com/go/warranty_terms• Three years for th
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 157Measurement CategoryThe Agilent U1251B and U1252B Handheld Digital Multimeter has a
158 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsSpecification Assumptions• The DC specifications are defined for measurements which
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 159Function Range ResolutionTest Current/ Burden VoltageAccuracyU1251B U1252BResistance
X Agilent U1251B/U1252B User’s and Service Guide Figure 3-8 Full Log 64Figure 3-9 Interval (Automatic) logging mode operation 66Figure 3-10 Log review
160 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsFunction Range ResolutionTest Current/ Burden VoltageAccuracyU1251B U1252BCurrent500
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 161AC SpecificationsAC specifications for U1251BTable 7-2 U1251B accuracy specification
162 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsAC specifications for U1252B Table 7-4 U1252B accuracy specifications ± (% of readin
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 163AC+DC Specifications for U1252BTable 7-6 U1252B true RMS ac+dc voltage specification
164 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsCapacitance SpecificationsTemperature SpecificationsTable 7-8 Capacitance specificat
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 165Frequency SpecificationsDuty Cycle and Pulse Width SpecificationsNotes for temperatu
166 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsFrequency Sensitivity SpecificationsFor voltage measurementsNotes for duty cycle spe
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 167For current measurements500.00 V 25 V - 30 V < 100 kHz- 50 V - 60 V -1000.0 V 50
168 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsPeak Hold SpecificationsFrequency Counter Specifications for U1252BTable 7-14 Peak h
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 169Square Wave Output for U1252BTable 7-16 Frequency counter (divide by 100 [4]) specif
Agilent U1251B/U1252B User’s and Service Guide XI List of TablesTa b l e 1 - 1 Rotary switch description and functions 8Ta b l e 1 - 2 Keypad descript
170 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsNotes for square wave output specifications:1 Output impedance: 3.5 kW maximum.2 The
Specifications 7Agilent U1251B/U1252B User’s and Service Guide 171Operating SpecificationsDisplay update rate (approximate)Table 7-18 Display update r
172 Agilent U1251B/U1252B User’s and Service Guide7 SpecificationsInput impedanceTable 7-19 Input ImpedanceFunction Range Input ImpedanceDC Voltage [1
© Agilent Technologies, Inc. 2009–2013Printed in MalaysiaTenth Edition, August 12, 2013U1251-90036www.agilent.comAgilent TechnologiesContact usTo obta
II Agilent U1251B/U1252B User’s and Service GuideNotices© Agilent Technologies, Inc. 2009–2013No part of this manual may be reproduced in any form or
XII Agilent U1251B/U1252B User’s and Service Guide Ta b l e 7 - 5 U1252B accuracy specifications ± (% of reading + num-ber of LSD) for true RMS AC cur
1Agilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service GuideAgilent Technologies1Getting StartedIntroducing the U1251B/U1252B Handhe
2 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialIntroducing the U1251B/U1252B Handheld Digital MultimeterKey features of thi
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 3Check the shipmentVerify that you have received the following items with yo
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 4Adjusting the tilt-standTo adjust the meter to a 60° standing position, pu
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 5To adjust the meter to a hanging position follow the steps shown in Figure
6 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialThe front panel at a glance Figure 1-4 U1252B front panelDisplayKeypadRotary
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 7The rear panel at a glanceFigure 1-5 Rear panelIR Communication PortTest pr
8 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialThe rotary switch at a glanceFigure 1-6 Rotary switchTable 1-1 Rotary switch
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 9The keypad at a glanceThe operation of each key is shown below. Pressing a
Agilent U1251B/U1252B User’s and Service Guide IIISafety SymbolsThe following symbols on the instrument and in the documentation indicate precautions
10 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started Tutorial5 scrolls through available measuring ranges (except when rotary switch is
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 11The display at a glanceTo view the full display (with all segments illumin
12 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialTable 1-3 General display symbolsLCD symbol DescriptionRemote control Therm
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 13The primary display signs are explained below. Table 1-4 Primary display s
14 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialThe secondary display signs are explained below.The analog bar emulates the
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 15The “+” or “–” sign is indicated when the positive or negative value has b
16 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started Tutorialdisplay and frequency on the primary display. Alternatively, pulse width (m
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 17Selecting display with the Dual buttonPress to select different combinat
18 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialTable 1-8 Selecting display with the Dual buttonRotary switch position (Fun
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 19(AC + DC voltage)[for U1252B]AC + DC mV Hz (AC coupling)dBm or dBV AC + D
IV Agilent U1251B/U1252B User’s and Service GuideSafety InformationThis meter is safety-certified in compliance with EN/IEC 61010-1:2001, UL 61010-1 S
20 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialSelecting display with the Shift buttonThe table below shows selection of p
Getting Started Tutorial 1Agilent U1251B/U1252B User’s and Service Guide 21 for U1252B(AC + DC Voltage)DC mVAC mVAC + DC mV Ω(Resistance)Ω ΩnS(Diode
22 Agilent U1251B/U1252B User’s and Service Guide1 Getting Started TutorialThe terminals at a glanceFigure 1-9 Connector terminalsWARNINGTo avoid dama
23Agilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service GuideAgilent Technologies2 Making MeasurementsUnderstanding The Measurement
24 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsUnderstanding The Measurement InstructionsWhen making measurements, follow the
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 25Measuring AC voltageSet up the multimeter to measure AC voltage as shown in
26 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsMeasuring DC voltageSet up the multimeter to measure DC voltage as shown in Fi
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 27
28 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsMeasuring AC and DC Signals (U1252B only)For better accuracy when measuring th
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 29Measuring CurrentµA & mA measurementSet up the multimeter to measure µA
Agilent U1251B/U1252B User’s and Service Guide VWARNING• When working above 70 VDC, 33 VAC RMS or 46.7 V peak, exercise caution – such range pose a sh
30 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFigure 2-3 Measuring μA and mA currentRed LeadBlack Lead
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 31Percentage scale of 4 mA to 20 mASet up the multimeter to measure percentage
32 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFigure 2-4 Measuring scale of 4-20 mARed LeadBlack Lead
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 33A (ampere) measurementSet up the multimeter to measure A (ampere) as shown i
34 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFrequency CounterSet up the multimeter to measure frequency as shown in Figure
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 35Figure 2-6 Measuring fre-quencyAudible continuityPressPressPress
36 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsMeasuring Resistance, Conductance and Testing ContinuitySet up the multimeter
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 37Press to scroll through audible continuity, conductance and resistance tes
38 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsAudible continuityIn the range of 0–500 Ω, the beeper will sound if the resist
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 39Figure 2-9 Conductance measurementRed LeadBlack Lead
VI Agilent U1251B/U1252B User’s and Service GuideRegulatory MarkingsThe CE mark is a registered trademarkof the European Community. This CEmark shows
40 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsTestin g DiodesTo test a diode, turn the power off to the circuit and remove t
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 41Figure 2-10 Measuring forward bias of diodeRed LeadBlack Lead
42 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFigure 2-11 Measuring reverse bias of diode Red LeadBlack LeadRed LeadBlack Le
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 43Measuring CapacitanceThe meter measures capacitance by charging the capacito
44 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFigure 2-12 Capacitance measurements
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 45Measuring TemperatureThe bead type thermocouple probe is suitable for making
46 Agilent U1251B/U1252B User’s and Service Guide2 Making Measurements8 For a quick measurement, use the 0 °C compensation adapter to see the temper
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 47If you are working in a constantly varying environment, where ambient temper
48 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsFigure 2-15 Surface temperature measurement
Making Measurements 2Agilent U1251B/U1252B User’s and Service Guide 49Alerts and Warning During Measurement Overload alertThe meter provides an ov
Agilent U1251B/U1252B User’s and Service Guide VIIWaste Electrical and Electronic Equipment (WEEE) Directive 2002/96/ECThis instrument complies with t
50 Agilent U1251B/U1252B User’s and Service Guide2 Making MeasurementsCharge terminal alertThe meter sounds an alert beep when the terminal detect
51Agilent TechnologiesAgilent U1251B and U1252B Handheld Digital Multimeter User’s and Service Guide3Functions and FeaturesDynamic Recording 52Data Ho
52 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsDynamic RecordingThe Dynamic Recording mode can be used to detectintermittent
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 53Figure 3-1 Dynamic recording mode operation Press Max Min for > 1
54 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsData Hold (Trigger Hold)The data hold function allows the user to freeze the
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 55Refresh HoldThe Refresh Hold function allows you to hold the displaye
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 56Figure 3-3 Refresh hold mode operationThe HOLD sign flashes until the
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 57Null (Relative)The Null function subtracts a stored value from the pr
58 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-4 Null (relative) mode operationPress Press as NULL is flashing on
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 59Decibel DisplayThe dBm operation calculates the power delivered to a
VIII Agilent U1251B/U1252B User’s and Service GuideIn This Guide…1Getting StartedThis chapter contains information on the Agilent U1251B and U1252B ha
60 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-5 dBm/dBV display mode operationPress PressPress Press Press for &
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 611 ms Peak HoldThe Peak Hold function allows the measurement of peak v
62 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-6 1 ms peak hold mode operationPress for > 1 sec.Press Press St
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 63Data LoggingThe data logging function provides the convenience of rec
64 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-7 Hand (Manual) logging mode operationFigure 3-8 Full LogPress for
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 65Interval loggingFirstly, ensure that interval (time) logging is speci
66 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-9 Interval (Automatic) logging mode operationPress for > 1 sec.P
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 67Reviewing logged data1 Press for more than 1 second to enter the Lo
68 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-10 Log review mode operationPress View for > 1 sec.Press ViewPres
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 69Square Wave Output (for U1252B)The square wave output function can be
Agilent U1251B/U1252B User’s and Service Guide IXDeclaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available o
70 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-11 Frequency adjustment for square wave output5 Press to select the
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 71Figure 3-12 Duty cycle adjustment for square wave outputPress Press P
72 Agilent U1251B/U1252B User’s and Service Guide3 Features and FunctionsFigure 3-13 Pulse width adjustment for square wave outputPress Press Press Pr
Features and Functions 3Agilent U1251B/U1252B User’s and Service Guide 73Remote CommunicationThe meter has a bi- directional (full duplex) com
74 Agilent U1251B/U1252B User’s and Service Guide3 Features and Functions
75Agilent TechnologiesAgilent U1251B and U1252B Handheld Digital MultimeterUser’s and Service Guide4Changing The Default SettingSelecting Setup Mode 7
76 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingSelecting Setup ModeTo enter the Setup mode, perform the following step
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 77Table 4-1 Available setting options in Setup modeMenu item Available se
78 Agilent U1251B/U1252B User’s and Service Guide4 Changing The Default SettingMenu item Available setting optionsDefault factory settingDisplay Descr
Changing The Default Setting 4Agilent U1251B/U1252B User’s and Service Guide 79Menu item Available setting optionsDefault factory settingDisplay Descr