Agilent Technologies
Agilent U1731C, U1732C,
and U1733C Handheld
LCR Meter
User’s Guide
Zobrazit stránku 0
1 2 3 4 5 6 ... 106 107

Shrnutí obsahu

Strany 1 - LCR Meter

Agilent TechnologiesAgilent U1731C, U1732C, and U1733C Handheld LCR MeterUser’s Guide

Strany 2

X U1731C/U1732C/U1733C User’s GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 3 - Safety Symbols

4 Characteristics and Specifications Electrical Specifications82 U1731C/U1732C/U1733C User’s GuideDissipation/Quality factor specificationsTable 4 -

Strany 4 - Safety Considerations

Characteristics and Specifications 4 Electrical SpecificationsU1731C/U1732C/U1733C User’s Guide 83Test signal specificationsTable 4 - 6 Test signal sp

Strany 5

4 Characteristics and Specifications Electrical Specifications84 U1731C/U1732C/U1733C User’s GuideSource impedance of impedance/resistance measuremen

Strany 6 - Environmental Conditions

Characteristics and Specifications 4 Electrical SpecificationsU1731C/U1732C/U1733C User’s Guide 85Source impedance of capacitance measurementTable 4 -

Strany 7 - Regulatory Markings

4 Characteristics and Specifications Electrical Specifications86 U1731C/U1732C/U1733C User’s GuideSource impedance of inductance measurementTable 4 -

Strany 8

Characteristics and Specifications 4 SMD Tweezers SpecificationsU1731C/U1732C/U1733C User’s Guide 87SMD Tweezers SpecificationsThe Agilent U1782B is a

Strany 9

4 Characteristics and Specifications SMD Tweezers Specifications88 U1731C/U1732C/U1733C User’s GuideElectrical characteristicsTable 5 U1782B SMD twee

Strany 10

© Agilent Technologies, Inc., 2011–2012Seventh Edition, November 28, 2012U1731-90077www.agilent.comAgilent TechnologiesContact usTo obtain service, wa

Strany 11 - Table of Contents

U1731C/U1732C/U1733C User’s Guide XITable of Contents1 IntroductionAbout This Manual 2Documentation map 2Safety notes 2Preparing Your L

Strany 12 - 3 Setup Options

XII U1731C/U1732C/U1733C User’s GuideMeasuring dissipation factor/quality factor/phase angle (D/Q/θ) 37Changing the test frequency 37Selecti

Strany 13

U1731C/U1732C/U1733C User’s Guide XIII4 Characteristics and SpecificationsProduct Characteristics 76Specification Assumptions 77Electrical S

Strany 14

XIV U1731C/U1732C/U1733C User’s Guide

Strany 15 - List of Figures

U1731C/U1732C/U1733C User’s Guide XVList of FiguresFigure 1-1 Installing the batteries 4Figure 1-2 Power-on button 5Figure 1-3 Tilt-stand adj

Strany 16

XVI U1731C/U1732C/U1733C User’s Guidemeasurements 60Figure 3-5 Changing the power-on secondary parameter and measurement mode for resistance (R) m

Strany 17 - List of Tables

U1731C/U1732C/U1733C User’s Guide XVIIList of TablesTab l e 1- 1 Battery level indicator 5Tab l e 1- 2 Power-on options 10Tab l e 1- 3 Fro

Strany 18

XVIII U1731C/U1732C/U1733C User’s Guide

Strany 19 - Introduction

1U1731C/U1732C/U1733C Handheld LCR MeterUser’s GuideAgilent Technologies1IntroductionAbout This Manual 2Documentation map 2Safety notes 2Prep

Strany 20 - About This Manual

II U1731C/U1732C/U1733C User’s GuideNotices© Agilent Technologies, Inc. 2011–2012No part of this manual may be reproduced in any form or by any means

Strany 21 - Preparing Your LCR Meter

1 Introduction About This Manual2 U1731C/U1732C/U1733C User’s GuideAbout This ManualThe descriptions and instructions in this manual apply to the Agi

Strany 22

Introduction 1 Preparing Your LCR MeterU1731C/U1732C/U1733C User’s Guide 3Preparing Your LCR MeterCheck the shipmentWhen you receive your LCR meter, c

Strany 23 - Turn on your LCR meter

1 Introduction Preparing Your LCR Meter4 U1731C/U1732C/U1733C User’s Guide1 Open the battery cover. Lift the tilt stand. Loosen the screw with a suit

Strany 24 - Enabling the backlight

Introduction 1 Preparing Your LCR MeterU1731C/U1732C/U1733C User’s Guide 5Turn on your LCR meterTo power ON your LCR meter, press the power- on button

Strany 25 - Selecting the range

1 Introduction Preparing Your LCR Meter6 U1731C/U1732C/U1733C User’s GuideTo power OFF your LCR meter, press the power- on button again. Automatic Po

Strany 26 - Adjusting the tilt stand

Introduction 1 Preparing Your LCR MeterU1731C/U1732C/U1733C User’s Guide 7Selecting the rangePressing switches the LCR meter between manual and auto

Strany 27 - Connecting the IR-USB cable

1 Introduction Preparing Your LCR Meter8 U1731C/U1732C/U1733C User’s GuideAdjusting the tilt standTo adjust the LCR meter to a 60° standing position,

Strany 28 - Power-on options

Introduction 1 Preparing Your LCR MeterU1731C/U1732C/U1733C User’s Guide 9Connecting the IR-USB cableYou can use the IR communication link (IR communi

Strany 29 - Your LCR Meter in Brief

1 Introduction Preparing Your LCR Meter10 U1731C/U1732C/U1733C User’s GuidePower-on optionsSome options can be selected only while you turn the LCR m

Strany 30 - Rear and side view

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 11Your LCR Meter in BriefDimensionsFront viewFigure 1-5 Width dimensions87 mm

Strany 31 - Overview

U1731C/U1732C/U1733C User’s Guide IIISafety SymbolsThe following symbols on the instrument and in the documentation indicate precautions which must be

Strany 32 - Rear panel

1 Introduction Your LCR Meter in Brief12 U1731C/U1732C/U1733C User’s GuideRear and side viewFigure 1-6 Height and depth dimensions184 mm41 mm

Strany 33

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 13OverviewFront panelThe front panel parts of your LCR meter are described in

Strany 34 - 1 Introduction

1 Introduction Your LCR Meter in Brief14 U1731C/U1732C/U1733C User’s GuideRear panelThe rear panel parts of your LCR meter are described in this sect

Strany 35 - Introduction 1

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 15KeypadThe operation of each key is explained below. Pressing a key enables a

Strany 36 - Display screen

1 Introduction Your LCR Meter in Brief16 U1731C/U1732C/U1733C User’s GuideSwitches between the dissipation factor (D), quality factor (Q), or phase a

Strany 37

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 17Sets the null/relative mode.• The displayed value is saved as a reference to

Strany 38

1 Introduction Your LCR Meter in Brief18 U1731C/U1732C/U1733C User’s GuideDisplay screenThe function that each display annunciator of your LCR meter

Strany 39 - Measurement units

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 19Tolerance mode indicator for sorting L, C, or Rpage 38Tolerance set to 1% fo

Strany 40 - Input terminals

1 Introduction Your LCR Meter in Brief20 U1731C/U1732C/U1733C User’s GuidePresent reading shown on primary displaypage 40Maximum reading shown on pri

Strany 41 - Cleaning Your LCR Meter

Introduction 1 Your LCR Meter in BriefU1731C/U1732C/U1733C User’s Guide 21Measurement unitsThe available signs and notations for each measurement func

Strany 42

IV U1731C/U1732C/U1733C User’s GuideSafety ConsiderationsRead the information below before using this instrument. The following general safety precaut

Strany 43 - Features and Functions

1 Introduction Your LCR Meter in Brief22 U1731C/U1732C/U1733C User’s GuideInput terminalsThe terminal and socket connections of your LCR meter are de

Strany 44 - Making Measurements

Introduction 1 Cleaning Your LCR MeterU1731C/U1732C/U1733C User’s Guide 23Cleaning Your LCR MeterDirt or moisture in the terminals can distort reading

Strany 45

1 Introduction Cleaning Your LCR Meter24 U1731C/U1732C/U1733C User’s GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 46 - 2 Features and Functions

25U1731C/U1732C/U1733C Handheld LCR MeterUser’s GuideAgilent Technologies2 Features and FunctionsMaking Measurements 26Auto Identification (Ai) fu

Strany 47 - P S

2 Features and Functions Making Measurements26 U1731C/U1732C/U1733C User’s GuideMaking MeasurementsAuto Identification (Ai) functionPress to automa

Strany 48

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 27The measurement mode (series or parallel) will be automatically identi

Strany 49

2 Features and Functions Making Measurements28 U1731C/U1732C/U1733C User’s GuideTable 2 - 3 Auto identification series/parallel rules for capacitance

Strany 50

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 29Measuring inductance (L)Set up your LCR meter to measure inductance as

Strany 51 - Measuring resistance (R)

2 Features and Functions Making Measurements30 U1731C/U1732C/U1733C User’s GuideoFigure 2-3 Measuring inductanceInductor32ORSurface Mount Device (SMD

Strany 52

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 31Measuring capacitance (C)Set up your LCR meter to measure capacitance

Strany 53 - Measuring impedance (Z)

U1731C/U1732C/U1733C User’s Guide VWARNING• Use this meter only as specified in this manual; otherwise, the protection provided by the meter may be im

Strany 54

2 Features and Functions Making Measurements32 U1731C/U1732C/U1733C User’s GuideFigure 2-5 Measuring capacitance32ORSurface Mount Device (SMD)U1782BS

Strany 55 - Changing the test frequency

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 33Measuring resistance (R)Set up your LCR meter to measure resistance as

Strany 56

2 Features and Functions Making Measurements34 U1731C/U1732C/U1733C User’s GuideFigure 2-7 Measuring resistanceResistor32ORSurface Mount Device (SMD)

Strany 57 - Enabling DCR measurements

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 35Measuring impedance (Z)All circuit components, resistors, capacitors,

Strany 58 - Additional Features

2 Features and Functions Making Measurements36 U1731C/U1732C/U1733C User’s Guide4 Press to change the secondary display measurement (D, Q, or θ).5

Strany 59 - > 1 s

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 37Measuring dissipation factor/quality factor/phase angle (D/Q/θ)The dis

Strany 60

2 Features and Functions Making Measurements38 U1731C/U1732C/U1733C User’s GuideSetting the standard reference tolerance (Tol%)The tolerance ranges a

Strany 61

Features and Functions 2 Making MeasurementsU1731C/U1732C/U1733C User’s Guide 39Enabling ESR measurementsPress for more than 1 second to select the

Strany 62

2 Features and Functions Additional Features40 U1731C/U1732C/U1733C User’s GuideAdditional FeaturesFreezing the display (Hold)To freeze the display f

Strany 63

Features and Functions 2 Additional FeaturesU1731C/U1732C/U1733C User’s Guide 41When the input goes below the recorded minimum value or above the reco

Strany 64 - Performing the open/short CAL

VI U1731C/U1732C/U1733C User’s GuideEnvironmental ConditionsThis instrument is designed for indoor use and in an area with low condensation. The table

Strany 65

2 Features and Functions Additional Features42 U1731C/U1732C/U1733C User’s GuideSetting the high/low limit comparison (Limit)The high and low limit c

Strany 66 - OPn) or short

Features and Functions 2 Additional FeaturesU1731C/U1732C/U1733C User’s Guide 43Press the key for more than 1 second to activate the high/low limit

Strany 67 - Features and Functions 2

2 Features and Functions Additional Features44 U1731C/U1732C/U1733C User’s GuideWhile the annunciator is flashing, use the or key to select an

Strany 68

Features and Functions 2 Additional FeaturesU1731C/U1732C/U1733C User’s Guide 45Press and hold for more than 1 second to exit this mode.Making relat

Strany 69 - Setup Options

2 Features and Functions Additional Features46 U1731C/U1732C/U1733C User’s GuideThe annunciator is shown on the display while the relative mode is

Strany 70 - Using the Setup Menu

Features and Functions 2 Additional FeaturesU1731C/U1732C/U1733C User’s Guide 47The open/short CAL function is available for fixed measurement ranges.

Strany 71 - Editing numerical values

2 Features and Functions Additional Features48 U1731C/U1732C/U1733C User’s GuideFigure 2-20 Open calibration and short calibration prompts1 Press and

Strany 72 - Setup Menu Summary

Features and Functions 2 Additional FeaturesU1731C/U1732C/U1733C User’s Guide 49THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Strany 73

2 Features and Functions Additional Features50 U1731C/U1732C/U1733C User’s Guide

Strany 74 - Setup Menu Items

51U1731C/U1732C/U1733C Handheld LCR MeterUser’s GuideAgilent Technologies3 Setup OptionsUsing the Setup Menu 52Editing numerical values 53Setup

Strany 75

U1731C/U1732C/U1733C User’s Guide VIIRegulatory MarkingsThe CE mark is a registered trademarkof the European Community. This CEmark shows that the pro

Strany 76 - Pr e s s

3Setup Options Using the Setup Menu52 U1731C/U1732C/U1733C User’s GuideUsing the Setup MenuThe Setup menu allows you to change a number of nonvolatil

Strany 77

Setup Options 3 Using the Setup MenuU1731C/U1732C/U1733C User’s Guide 53Editing numerical valuesWhen editing numerical values, use the and to posi

Strany 78

3Setup Options Setup Menu Summary54 U1731C/U1732C/U1733C User’s GuideSetup Menu SummaryThe Setup menu items are summarized in the table below. Click

Strany 79

Setup Options 3 Setup Menu SummaryU1731C/U1732C/U1733C User’s Guide 55H01 to H16 or L01 to L160 to 19999Set the high and low limits for each variable

Strany 80

3Setup Options Setup Menu Items56 U1731C/U1732C/U1733C User’s GuideSetup Menu ItemsChanging the initial power-on behaviorYou can change the power- on

Strany 81

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 57The LCR meter will start up in the selected measurement type for subsequent power

Strany 82 - 3Setup Options

3Setup Options Setup Menu Items58 U1731C/U1732C/U1733C User’s GuideChanging the power-on test frequencyUse this Setup item to change the LCR meter’s

Strany 83 - • User (Ur01 to Ur16)

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 59Changing the power-on secondary parameter and measurement mode for inductance (L)

Strany 84

3Setup Options Setup Menu Items60 U1731C/U1732C/U1733C User’s GuideChanging the power-on secondary parameter and measurement mode for capacitance (C)

Strany 85 - Setup Options 3

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 61Changing the power-on secondary parameter and measurement mode for resistance (R)

Strany 86 - Changing the baud rate

VIII U1731C/U1732C/U1733C User’s GuideWaste Electrical and Electronic Equipment (WEEE) Directive 2002/96/ECThis instrument complies with the WEEE Dire

Strany 87 - Changing the parity check

3Setup Options Setup Menu Items62 U1731C/U1732C/U1733C User’s GuideChanging the power-on open/short CALUse this Setup item to change the LCR meter’s

Strany 88 - Changing the data bits

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 63Changing the Ai function’s phase angle conditionThis setting is used with the Ai f

Strany 89

3Setup Options Setup Menu Items64 U1731C/U1732C/U1733C User’s GuideFigure 3-7 Changing the Ai function’s phase angle conditionPr e s sPr e s sPr e s

Strany 90 - Locking the push buttons

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 65Changing the power-on limit category and setThis setting is used with the Limit co

Strany 91

3Setup Options Setup Menu Items66 U1731C/U1732C/U1733C User’s GuideChanging the user high/low limit valuesThis setting is used with the Limit compari

Strany 92 - Press

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 67Figure 3-9 Changing the user high/low limit valuesU11 6800 6120U12 8200 7380U13 10

Strany 93

3Setup Options Setup Menu Items68 U1731C/U1732C/U1733C User’s GuideChanging the baud rateThis setting is used with the IR communication link and the

Strany 94 - Product Characteristics

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 69Changing the parity checkThis setting is used with the IR communication link and t

Strany 95 - Specification Assumptions

3Setup Options Setup Menu Items70 U1731C/U1732C/U1733C User’s GuideChanging the data bitsThis setting is used with the IR communication link and the

Strany 96

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 71Changing the beep frequencyThe LCR meter’s beeper alerts users to the presence of

Strany 97 - Capacitance specifications

U1731C/U1732C/U1733C User’s Guide IXDeclaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available on the Agilent

Strany 98 - Inductance specifications

3Setup Options Setup Menu Items72 U1731C/U1732C/U1733C User’s GuideLocking the push buttonsUse this Setup item to lock the push buttons (keys) of you

Strany 99

Setup Options 3 Setup Menu ItemsU1731C/U1732C/U1733C User’s Guide 73Changing the auto power-off and backlight time-outsThe LCR meter’s automatic power

Strany 100 - Electrical Specifications

3Setup Options Setup Menu Items74 U1731C/U1732C/U1733C User’s GuideResetting the Setup itemsThe Setup items can be reset to their default values thro

Strany 101 - Test signal specifications

75U1731C/U1732C/U1733C Handheld LCR MeterUser’s GuideAgilent Technologies4Characteristics and SpecificationsProduct Characteristics 76Specification

Strany 102

4 Characteristics and Specifications Product Characteristics76 U1731C/U1732C/U1733C User’s GuideProduct CharacteristicsNOTEProduct characteristics sp

Strany 103

Characteristics and Specifications 4 Specification AssumptionsU1731C/U1732C/U1733C User’s Guide 77Specification Assumptions• Accuracy is given as ±(%

Strany 104

4 Characteristics and Specifications Electrical Specifications78 U1731C/U1732C/U1733C User’s GuideElectrical SpecificationsImpedance/Resistance/DCR s

Strany 105 - SMD Tweezers Specifications

Characteristics and Specifications 4 Electrical SpecificationsU1731C/U1732C/U1733C User’s Guide 79Capacitance specificationsTable 4 - 2 Capacitance sp

Strany 106 - Electrical characteristics

4 Characteristics and Specifications Electrical Specifications80 U1731C/U1732C/U1733C User’s GuideInductance specificationsTable 4 - 3 Inductance spe

Strany 107 - Agilent Technologies

Characteristics and Specifications 4 Electrical SpecificationsU1731C/U1732C/U1733C User’s Guide 81Phase angle of impedance specificationsTable 4 - 4 P

Komentáře k této Příručce