1100 LC Modules Control Manual - 164 - 03-914947-30:Rev.6
The Limit values correspond to the maximum allowed values for
the counters before the maintenance is needed.
The Actual column displays the incremented counter value since
the last reset of the counters.
In the Limit and Actual columns, select the current value and
modify it. Validate by pressing the
button.
The counter values of the VIS lamp Ignition, VIS lamp Burn, UV
lamp Ignition and UV lamp Burn cannot be modified. The Actual
values are respectively equal to the Lamp Ignition and Lamp On-
Time values displayed in the detector status.
The Accumulated values correspond to counter value since the
first operation of the pump module.
At the right of the table, a visual illustration shows the usage of
each counter. Below 75%, the usage is typified in green and
above 75% in red to alert that maintenance will be required soon.