Agilent Technologies 1100 Series Manuál s instrukcemi Strana 116

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 272
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 115
116
Repairing the Pump
Installing the Continuous Seal Wash Option
Installing the Continuous Seal Wash Option
Tools required:
4-mm hexagonal key
Parts:
Seal wash kit (01018-68722)
1 Disconnect all capillaries and tubes from the
pump head and disconnect the active inlet
valve cable.
2 Using a 4-mm hexagonal key stepwise
loosen and remove the two pump head
screws and remove the pump head from the
pump drive.
3 Place the pump head on a flat surface. Loosen
the lock screw (two revolutions) and while
holding the lower half of the assembly
carefully pull the pump head away from the
plunger housing.
Pump head screws
Pump head
Lock screw
Zobrazit stránku 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 271 272

Komentáře k této Příručce

Žádné komentáře