Audio Frequency Commands (Measure Subsystem)
ACLevel
ACLevel
Description Sets the AC Level MEASurement attributes.
GPIB unit is V.
Display units are dBm, V, mV, uV, dBuV, W; default unit is V.
Queries the AC Level MEASurement result for AF Analyzer inputs.
(AFAN:INP) that are in units of AC level.
Syntax MEASure:AFRequency:ACLevel?
MEASure:AFRequency:ACLevel[:MM] | [:AVG] | [:MET]
Options Refer to Appendices D, F and G.
AM
Description Sets the AM Depth MEASurement attributes.Queries the AM Depth MEASurement result
for AF Analyzer inputs (AFAN:INP) that are units of percent.
GPIB unit is %(PCT).
Display units are %(PCT).
Syntax MEASure:AFRequency:AM?
MEASure:AFRequency:AM[:MM] | [:AVG] | [:MET]
Options Refer to Appendices D, F and G.
DCAM
Description Sets the DC AM Level MEASurement attributes. Queries the DC Level MEASurement
result for AF Analyzer inputs (AFAN:INP) that are units of percent.
GPIB unit is %(PCT).
Display units are %(PCT).
Syntax MEASure:AFRequency:DCAM?
MEASure:AFRequency:DCAM[:MM] | [:AVG] | [:MET]
Options Refer to Appendices D, F and G.