To select TV display grid
• Press Display , then press the Grid softkey until TV is selected.
To autoscale on a video signal
1 Use a cable to connect a TV signal to channel 1.
2 Press Mode in the TRIGGER section of the front panel, and
select the
Trigger Mode TV softkey.
3 To select a TV standard, press Slope/Coupling in the TRIGGER
section of the front panel, then press the Standard softkey to select
the TV standard. Your cho
ices are NTSC, PAL, SECAM, and GENERIC.
GENERIC is used for other TV/Video standards. If your TV standard
has been previously selected, you may skip this step.
Use NTSC Instead of PAL-M
To trigger on a PAL-M signal, use NTSC. The line and field rates are identical.
4
Press Mode , then press the Video Autoscale softkey.
Provide correct source matching
Many TV signals are produced from 75Ω sources. To provide correct matching
to these sources, an 11094B 75Ω load is included as an accessory. For
oscilloscopes that have selectable input impedance, the 1 MΩ input should be
used with the 75Ω load.
The Undo Autoscale softkey in the
Setup menu resets the instrument
to the configuration that was present before pressing Video Autoscale.
Using Option 005 Enhanced TV/Video Trigger
To select TV display grid
3–4