Agilent Technologies 71501D Uživatelský manuál Strana 45

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 176
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 44
2-11
Application Tutorials
Tutorial 3: Measure Extinction Ratios on Low-Level Signals
Autoscaled Display of Low-Level Signal
4 Turn filtering on by pressing:
Setup, eyefltr ON OFF ON
Perform an Offset Calibration
This calibration procedure removes any offset that may be present in the opti-
cal-to-electrical converter. This is sometimes referred to as the "dark" level.
The offset calibration ensures accurate measurements of the lasers level.
5 Turn the laser off. (If you are measuring the pattern generator directly,
disconnect the input signal at channel 1.)
6 Perform the calibration by pressing:
Trg,Cal, OFFSET CAL, CONT
The calibration takes about a minute to execute. When the calibration is
finished, DC NULL: done is displayed.
7 Turn the laser on. (If you are measuring the pattern generator directly,
reconnect the input signal at channel 1.)
Zobrazit stránku 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 175 176

Komentáře k této Příručce

Žádné komentáře