Agilent Technologies E8267D PSG Uživatelský manuál Strana 6

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 24
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 5
6
20 GHz level accuracy
Resolution 0.01 dB
Temperature stability 0.01 dB/ °C (typical)
User flatness correction
Number of points 2 to 1601 points/table
Number of tables Up to 10,000, memory limited
Path loss Arbitrary, within attenuator range
Entry modes Remote power meter
1
, remote bus, manual
(user edit/view)
Output impedance 50 (nominal)
SWR (internally leveled, typical)
250 kHz to 2 GHz < 1.4:1
> 2 GHz to 20 GHz < 1.6:1
Leveling modes Internal leveling, external detector leveling,
millimeter source module, ALC Off
External detector leveling
Range –0.2 mV to –0.5 V, (nominal) (–36 dBm to +4 dBm
using Agilent 33330D/E detector)
Bandwidth 10 kHz (typical)
(Note: not intended for pulsed operation)
Maximum reverse power 1/2 Watt (nominal)
1. Compatible with Agilent Technologies EPM Series (E4418B and E4419B) power meters.
Measured level accuracy +5 dBm
Frequency (MHz)
1
0.5
0
-0.5
-1
0
5000
10000 15000
20000
Zobrazit stránku 5
1 2 3 4 5 6 7 8 9 10 11 ... 23 24