Agilent Technologies EXG Uživatelská příručka Strana 178

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 426
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 177
164 Agilent X-Series Signal Generators User’s Guide
Basic Digital Operation (Option 653/655/656/657)
Using Waveform Markers
Close–up of averaging
The ALC samples the waveform when the marker signal goes
high, and uses the average of the sampled waveform to set the
ALC circuitry.
Here the ALC samples during the on marker points (positive
polarity).
Marker
Marker
MarkerMarker
Example of Correct Use
Waveform: 1022 points
Marker range: 95–97
Marker polarity: Positive
This example shows a marker set to sample the waveform’s area of
highest amplitude. Note that the marker is set well before the
waveform’s area of lowest amplitude. This takes into account any
response difference between the marker and the waveform signal.
Pulse Unleveled
Example of Incorrect Use
Waveform: 1022 points
Marker range: 110–1022
Marker polarity: Positive
This example shows a marker set to sample the low part of the
same waveform, which sets the ALC modulator circuitry for
that level; this usually results in an unleveled condition for the
signal generator when it encounters the high amplitude of the
pulse.
MarkerMarker
Zobrazit stránku 177
1 2 ... 173 174 175 176 177 178 179 180 181 182 183 ... 425 426

Komentáře k této Příručce

Žádné komentáře