Instrument Operating System
Disk Drive Recovery Process
NOTE After the final reboot, you might see an error dialog box appear on the XSA
applications’s splash screen. This typically occurs when there is a significant
difference between the software version installed prior to performing the recovery
and the version restored by the recovery. For example, recovering from version
A.02.06 to A.01.57 will often cause this to occur.
Updating the software
The following steps load a copy of the instrument software that forces the program
code resident in the instrument’s various programmable gate arrays to be in sync
with the restored system software.
Removing the Software
Step Action
1. Connect a USB keyboard and mouse
to the instrument
2. Log out of the default user
(instrument):
•Select Start, Log Off, Log Off
3. At the log in prompt enter: a. User Name: administrator
b. Password: agilent4u
3. Uninstall the application: a. Select Start, Control Panel, Add
or Remove Programs
b. Find the Agilent X-Series Signal
Analyzer Application, and select
Remove.