Agilent Technologies Agilent G1369B Uživatelský manuál Strana 138

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 206
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 137
138 Agilent 1260 Infinity High Performance Autosampler User Manual
9 Maintenance
Installing the Metering Seal
Installing the Metering Seal
When When poor injection volume reproducibility or when metering device / analytical head is leaking.
Tools required p/n Description
8710-0510 Wrench open 1/4 — 5/16 inch
8710-2392 4 mm Hex key
G4226-43800 Seal insert tool
Parts required # p/n Description
1 5063-6589 Metering seal (pack of 2) for 100 µL analytical head
Preparations Removing the metering seal, see “Removing the Metering Seal” on page 135.
1 Install the new metering seal using the plastic side of the
insert tool. Press it firmly into position. Avoid any offset
angle as it might deform the seal.
2 Reassemble the metering device / analytical head. Make
sure to tighten screws firmly and have the tag on the
right side looking from the front.
>chZgiidda
Zobrazit stránku 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 205 206

Komentáře k této Příručce

Žádné komentáře