STATus
The STATus subsystem reports the bit values of the Operation Status Register.
Enables the Status Register to set a bit after a bit is set to 1 by the Operation Status
Register.
Subsystem Syntax STATus
:OPERation
:ENABle <
number>
[:EVENt]?
:OPERation:ENABle
STATus:OPERation:ENABle <number> enables the Operation Status Register to
set a bit in the Status Register. For multiplexer modules, when bit 8 in the Operation
Status Register is set to 1, bit 7 in the Status Register is set to 1.
Parameters
Comments Setting Bit 7 of the Status Register: STATus:OPERation:ENABle 256 sets bit
7 of the Status Register to 1 after bit 8 of the Operation Status Register is set to
1.
Related Commands: [ROUTe:]SCAN
Example Enable the Status Register
STAT:OPER:ENAB 256
Enables bit 7 of the Status Register.
Parameter Name Parameter Type Range of Values
<
number>
numeric 1–32768