Chapter 5: Replacing Assemblies
Removing and Replacing the Fans
102
Removing and Replacing the Fans
WARNING AVOID INJURY!
The fan blades are exposed both inside and outside the chassis. Disconnect the power
cable before working around the fan. Use extreme caution in working with the
oscilloscope. Failure to observe these precautions may result in injury.
1
Disconnect the power cable and remove the top panel, power supply, front
panel, bottom panel, and acquisition assembly as described earlier.
2 Remove the six M3 screws connecting the fan bracket to the chassis. When
replacing, torque the screws to 5 in- lbs.
Figure 5-67
3
Remove the fan assembly, carefully feeding the fan assembly cable through the
bottom of the chassis.
Figure 5-68