DisplayPort Test Implementation Group 26 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Note: Crosstalk is highly dependent on the Receptacle fixture used.
b. Alternate between adjusting the Vamp on the J-BERT N4903B user interface and pressing
the Single Acquisition Button on the Oscilloscope until the Eye Height is calibrated to the
following value.
For RBR
For HBR
For HBR2
NA
NA
Adjust the Clock/Trigger Vamp
Height measurement on the
Scope is 90mV
diff
+/-10%
Repeat measurement at least 5
times and average the results.
Figure 25: HBR2 Eye Height Measurement with crosstalk
2. At TP3, disconnect ‘P’ (Plug) type adapter from the ‘R’ (Receptacle) calibration adapter and connect
the ‘P’ type adapter to the DUT for testing (Figure 4-4).
a. At TP3, disconnect ‘P’ (Plug) type adapter from the ‘R’ (Receptacle) calibration adapter and
connect the ‘P’ type adapter to the DUT for testing (Figure 4-4).
b. Disconnect the Plug fixture from the calibration Receptacle fixture for the test. Refer to the
Testing the DUT section below.
Note: Always calibrate for a four lane configuration even if the device under test is a
single lane or two lane device. Perform the testing with aggressor signals applied on all
lanes independent of the actual device configuration.