434
Troubleshooting the Logic Analyzer
Error Messages
"Must have at least 1 edge specified"
You must assign at least one clock edge to one of the available clocks in
the clocking arrangement. The analyzer will not let you close the clock
assignment pop-up until an edge is specified.
"Time correlation of data is not possible"
To time-correlate data, the data must be stored with time tags.
Set the Count field in the Analyzer Trigger menu to Time.
"Maximum of 32 channels per label"
You have tried to assign more than 32 channels to a single label.
Unassign some of the channels. If you need more than 32 channels to
specify trigger conditions, you can AND terms in the Analyzer Trigger
menu.