The more keytoggles
beh*'een the two levels of the channel
menu.
probe
key
Channei Menu
5‐4
The probe
key
selects
probe
attenuation with a
range
of 1000:1
to .9000:L.
Attenuation
is
adjusted by either knob
or entry
keypad.
When
the knob is
in
coarse mode,
adjustments are
incremented
or decremented
in
the
1-2-5
sequence. When
in fine mode
adjustments
are in
0.1
increments.
Probe
attenuation
affects scaling
factors for the display,
not
sensitivity at
the input.
Probe
attenuation is
calibrated
in
the Utility
menu.
See
Chapter
L2,
"Utility
Menu",
for information
on
probe
calibraation.
Attenuation factors
are saved
with
the
front
panel
setup.
HP 54510A
Front‐
Pane:Reference