SCPI Commands Reference Status Handling Commands
78 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
:STATus:QUEStionable:MONotony:ENABle
Syntax
:STAT:QUES:MON:ENAB[?]
Form Set & Query
Description Sets or queries the enable register in the QUEStionable:MONotony
status group.
Parameter Numeric
*RST value
Specified limits 0 … 32767
:STATus:QUEStionable:MONotony:NTRansition
Syntax
:STAT:QUES:MON:NTR[?]
Form Set & Query
Description Sets or queries the negative- transition register in the
QUEStionable:MONotony status group.
NOTE The Agilent 81133A/81134A Pulse Generator does not use the
transition registers of the QUEStionable:MONotony status group,
therefore, this command is redundant.
Parameter Numeric
*RST value
Specified limits 0 … 32767
:STATus:QUEStionable:MONotony:PTRansition
Syntax
:STAT:QUES:MON:PTR[?]
Form Set & Query
Description This command sets or queries the positive- transition register in
the QUEStionable:MONotony status group.