Auxiliary Channels
Outputs
Trigger Output
The trigger output can be enabled/disabled. The levels of the trigger output can be set as high level - low level pair.
The trigger output can be set to one of the following modes:
· Trigger on clock
The frequency of the trigger output is set to the system frequency. It can be further divided by n (n= 1, 2, 3, 4, 5, 6,
7…2
31
-1).
· Trigger on data
One Trigger pulse occurs on every start of the repetitive data pattern.
Inputs
Note: The built in input termination eliminates the need for external bias networks and prevents a degrading of
the input sensitivity
Clock Input
The clock input can be ‘AC’ or ‘DC’ terminated. The ‘DC’ termination voltage can be set. See also ‘Clock Source -
external’.
Start Input
The start input can be used to start the instruments. When armed the instrument waits for the selected edge of
the applied signal.
Parameters:
· Threshold (voltage)
· Edge (rising/falling)
· Termination voltage
Store/Recall
Allows permanent storage of instrument settings. 9 store/recall registers are provided. Included in the stored
settings are all parameters and data.
Overprogramming
Many parameters can be programmed to values that exceed the specified ranges.