2-14 Agilent B1500 VXIplug&play Driver Users Guide, Edition 3
Driver Functions
Status Code
After measurement is performed, the Agilent B1500 returns a status code to notify
you if the measurement has been completed successfully. The status code will be
returned with the measurement data by the following functions that perform
measurement. Available status values are listed in Table 2-8.
“agb1500_spotCmuMeas”
“agb1500_spotMeas”
“agb1500_measureM”
“agb1500_measureP”
“agb1500_sampleIv”
“agb1500_sweepCv”
“agb1500_sweepIv”
“agb1500_sweepMiv”
“agb1500_msweepIv”
“agb1500_msweepMiv”
“agb1500_sweepPiv”
“agb1500_sweepPbias”
“agb1500_measureBdv”
“agb1500_measureIleak”
“agb1500_readData”
NOTE If multiple status conditions were found
Sum of the status values will be returned. For example, if an A/D converter
overflow occurred, and an SMU was oscillating during the measurements, the
returned value is 3 (=1+2).