Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 17
Classic Test Definition
I/V List Sweep
Constant This area defines the constant output source.
Unit SMU used for the constant source
V Name Variable name for the constant voltage output or measurement
data
I Name Variable name for the constant current output or measurement
data
Mode Output mode, V (voltage), I (current), VPULSE (voltage pulse),
or IPULSE (current pulse)
Source Output value
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Sweep status This field specifies the sweep abort condition. See “Sweep Abort Function” on page
4-37.
Range This button opens a window for setting the measurement ranging operation. See
“SMU Range Setup Window” on page 2-48.
ADC/Integ This button opens a window for setting the A/D converter of the measurement
channels. See
“ADC and Integration Time Setup Window” on page 2-50.
Advanced This button opens a window for setting other functions of the measurement channel.
See
“Advanced Setup Window” on page 2-53.
SPGU Pulse /
ALWG Setup
This button opens a window used to set the SPGU output. See “SPGU Pulse Setup
Window” on page 2-63 or “SPGU ALWG Setup Window” on page 2-67.