Agilent Technologies EasyEXPERT Uživatelská příručka Strana 270

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 344
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 269
4- 26 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Function Details
C-V Sweep Measurement
The MFCMU will internally perform the error correction for the CMU cables and
CMU accessories that keep the four-terminal pair configuration.
For the measurements for evaluating DUT against a working standard, perform the
open/short/load correction.
However, you have to consider that some parasitic remain in measurement path
even after performing corrections. To minimize the influence of parasitic, make the
signal path from the shield ends to the DUT as short as possible.
To perform the phase compensation and error correction, see “CMU Calibration” on
page 1-60.
Figure 4-14 Error Correction
KLJK
&08UHWXUQ
ORZ
9LUWXDOJURXQG
0XOWLIUHTXHQF\&08
9
+FXU
$
/FXU
/SRW
$XWREDODQFHEULGJH
6\VWHPJURXQG
9[
2XWHUVKLHOG
LQWHUQDO
H[WHUQDO2SHQ6KRUW/RDG
9[
'87
WHUPLQDOSDLUV
WHUPLQDO
(UURUFRUUHFWLRQ
&RQQHFWLRQ
(UURUHOHPHQWV
+SRW
Zobrazit stránku 269
1 2 ... 265 266 267 268 269 270 271 272 273 274 275 ... 343 344

Komentáře k této Příručce

Žádné komentáře