Log digitization uncertainty
10 dB/div ±0.2dB
5 dB/div ±O.I dB
2 dB/div ± 0.04 dB
1 dB/div ± 0.02 dB
Linear digitization uncertainty ± 0.2% of ref level
Error correction accuracy ± 0.4 dB
(applicable when SHIFT W
and SHIFT X are executed)
Reference line accuracy Equals the sum of reference level accuracy plus the scale fidelity
between the reference level and the reference line level.
Dynamic range
Spurious responses (signals generated by the analyzer due to input signals) for signals
< -40 dBm at the input mixer, all harmonic and intermodulation distortion > 70 dB
below input signal.
Second harmonic distortion (for mixer levels < -40 dBm)
100 Hz to 50 MHz < -70 dBc
50 to 700 MHz < -80 dBc
700 MHz to 2.5 GHz < -70 dBc
For mixer levels -10 dBm
2 to 22 GHz <-100 dBc
Third order intermodulation distortion
Third order intercept (TOI)
100 Hz to 5 MHz > +5 dBm
5 MHz to 5.8 GHz > +7 dBm
5.8 GHz to 18.6 GHz > +5 dBm
18.6 to 22 GHz > +5 dBm (typical)
2 to 22 GHz, for > 100 MHz > +50 dBm (typical) signal separation
Image, multiple, and out-of-band responses
Image responses are due to input signals that are two times the IF frequency above or
below the tuned frequency. Multiple responses are due to input signals mixing with more
than one LO harmonic. Out-of-band responses are due to input signals outside of the
selected frequency band.
Figure 4. Typical optimum dynamic range
Specifications (continued)