Chapter 5: Replacing Assemblies
To remove and replace the USB or GPIB port
125
To remove and replace the USB or GPIB port
Use this procedure to remove and replace the USB or GPIB port (GPIB is an option and may not
be on your oscilloscope). The procedure is exactly the same for both of these ports so only one
will be shown here. When necessary, refer to other removal procedures. The graphics in this
chapter are representative of the oscilloscope at the time of this printing. Your unit may look
different.
1
Disconnect the power cable and remove the cover and top plate.
2
Remove the screw from top edge of chassis (Figure 5-56).
Figure 5-56
3
Undo the connection to the motherboard by disconnecting the cable.
4
Pull the port out of the instrument.
5
To replace one of these ports, reverse the directions.