Chapter 3: Testing Performance
Performance Test Record
46
Offset Gain Test
Vertical
Sensitivity
V
Aux Out
Setting
V
DMM+
V
DMM-
V
DMM0
V
Scope+
V
Scope-
V
Scope0
Calculated
Offset Gain
Error
Offset Gain
Error Test
Limits
Channel 1
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 2
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 3
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 4
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %