PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 333
Reset Condition
On reset, the CCDF cummulative counts will be set to the default value,
100 M samples.
Query
PSTatistic[1]|2:CCDF:COUNt?
The query returns the current numeric value of the CCDF cummulative
count for the respective channel selected.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in free- run acquisition mode, error –221 "Settings
conflict" occurs.
If user inputs a count that is lesser than 100 M or greater than 10 G,
error –222 "Data out of range" occurs. Any input count that is within
range will be normalized and round down to the nearest 100 M.
PST1:CCDF:COUN? This command queries the numeric value
of CCDF cummulative counts for Channel
A.