Agilent Technologies 1912A Servisní příručka Strana 438

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 803
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 437
10 SENSe Subsystem
410 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
This command sets the measurement mode for the E9320 and N1920
sensor
either to average or peak sensor mode.
Syntax
Parameters
1
When measurement mode is set to average:
If TRIG:SOUR is set to INT1, INT2 or EXT, it is set automatically to IMM.
INIT:CONT is set automatically to ON.
SENS:AVER2:STAT is set automatically to OFF.
CALC:FEED is set automatically to “POW:AVG for all CALC blocks using the specified channel in
their CALC:MATH:EXPR.
2
When measurement mode is set to NORMal:
SENS:CORR:DCYC:STAT is set automatically to OFF.
Item Description/Default Range of Values
character_data Defines the measurement mode:
AVERage: sets the E9320 and P-Series
sensor to average only mode.
NORMal: sets the E9320 and P-Series
sensor to normal mode.
AVERage
1
NORMal
2
DET
:FUNC
SENS
1
SENS2
:
Space
?
character_data
Zobrazit stránku 437
1 2 ... 433 434 435 436 437 438 439 440 441 442 443 ... 802 803

Komentáře k této Příručce

Žádné komentáře