• Measure up to 1000 volts with61⁄2digits resolution• 0.0015% basic dcV accuracy (24 hour)• 0.06% basic acV accuracy (1 year)• 3Hz to 300kHz ac bandwi
Front-Panel Display Formats -H.DDD,DDD EFFFFront-panel display format. 10.216,5 VDCThis is the 10 Vdc range, 51⁄2 digits are displayed. -04
Resistance MeasurementsThe HP 34401A offers two methods for measuring resistance: 2-wireand 4-wire ohms. For both methods, the test current flows fro
True RMS AC Measurements True RMS responding multimeters, like the HP 34401A, measure the“heating” potential of an applied voltage. Unlike an “average
An ac-coupled true RMS measurement is desirable in situations whereyou are measuring small ac signals in the presence of large dc offsets.For example,
2Accuracy Specifications ± (% of reading + % of range)[1]Temperature Coefficient Frequency, 24 Hour [2]90 Day 1 Year 0°C – 18°CFunction Range[3]etc. 2
3Measurement Characteristicsdc VoltageMeasurement Method Continuously Integrating Multi-slope IIIA-D ConverterA-D Linearity 0.0002% of reading + 0.000
Ordering InformationAgilent 34401A MultimeterAccessories includedTest Lead Kit with probe, alligator, and grabber attachments, operatingmanual, servic
To Measure Voltage Ranges: 100 mV, 1 V, 10 V, 100 V, 1000 V (750 Vac)Maximum resolution: 100 nV (on 100 mV range)AC technique: true RMS, ac-
To Measure Current Ranges: 10 mA (dc only), 100 mA (dc only), 1 A , 3 AMaximum resolution: 10 nA (on 10 mA range)AC technique: true RMS, ac-co
To Test Continuity Test current source: 1 mAMaximum resolution: 0.1 Ω (range is fixed at 1 kohm)Beeper threshold: 1 Ω to 1000 Ω (beeps below ad
To Select a Range You can let the multimeter automatically select the range usingautoranging or you can select a fixed range using manual ranging. •
To Set the Resolution You can set the display resolution to 41⁄2, 51⁄2, or 61⁄2 digits either tooptimize measurement speed or noise rejection. In th