Agilent Technologies 53230A Uživatelská příručka Strana 85

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 294
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 84
53220A/53230A User’s Guide 85
53220A/53230A Measurements 3
Period
A standard period measurement is shown in Figure 3-4.
Figure 3-4. Standard Period Measurement.
The commands used to make period measurements are:
MEASure:PERiod? [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
CONFigure:PERiod [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
- expected is the expected input signal period. resolution is the desired
measurement resolution in seconds. The parameter ranges are:
expected (channels 1 and 2): 2.8 nsec to 10 sec (default = 100 nsec)
expected (channel 3 Option 106): 160 psec to 10 nsec (default = 2 nsec)
expected (channel 3 Option 115): 66 psec to 3.33 nsec (default = 2 psec)
+ width
-width
t
r
t
f
period
10
90
Zobrazit stránku 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 293 294

Komentáře k této Příručce

Žádné komentáře