Agilent Technologies 8935 Series Specifikace Strana 168

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 343
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 167
Chapter 7 167
Periodic Adjustments
Periodic Calibration Menu Descriptions
Figure 7-4 Measurement (A2A33) Assembly Test Points
Audio Frequency Generator Gain
The gain of the following paths is calibrated:
The internal paths that run from Audio Frequency Generators 1 and
2 (individually) through the Modulation Distribution assembly, to
the monitor select output, then onto Audio Analyzer 1 to the DVM.
The paths that run from Audio Frequency Generators 1 and 2
(individually) through the Modulation Distribution assembly to the
AUDIO OUT connector, externally to the rear-panel MODULATION
IN connector, then again through the Modulation Distribution
assembly to the monitor select output and to the DVM.
The above-measured levels are used to adjust the output level of the
audio generators so that they produce a calibrated level to the
modulation inputs of the RF generator. These measurements are made
at DC. Both positive and negative levels are measured to produce an
optimum calibration factor.
SGND
+REF -REF
Measurement Board Assembly
a2a33_1.eps
Zobrazit stránku 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 342 343

Komentáře k této Příručce

Žádné komentáře